Our days, light microscopy techniques providing high/super resolution imaging in fixed and living specimen are developing rapidly. Super-resolution techniques such as stimulated emission depletion (STED), structured illumination microscopy (SIM) and photo-activation localization microscopy (PALM) or direct stochastic optical reconstruction microscopy (STORM), respectively, are commercially available and enable imaging of subcellular structures showing details almost on the molecular scale1,2,3,4,5,6. However, these approaches still have limited applicability for live imaging experiments in which large volumes need to be visualized with multiple frames per second acquisition speed. Varieties of highly dynamic processes regulated via the plasma membrane, e.g. endo-/exocytosis, adhesion, migration or signaling, occur with high speed within large cellular volumes. Recently, in order to fill up this gap, an integrated microscopy technique was proposed called spinning disk-TIRF (SD-TIRF)7. In detail, TIRF microscopy permits to specifically isolate and localize the plasma membrane8,9, while SD microscopy is one of the most sensitive and fast live imaging techniques for the visualization and tracking of subcellular organelles in the cytoplasm10,11. The combination of both imaging techniques in a single setup has already been realized in the past12,13, however, the microscope presented here (Figure 1) finally meets the criteria to perform live imaging SD-TIRF experiments of the aforementioned processes at 3 frames per second speed. Since this microscope is commercially available, the goal of this manuscript is to describe in details and provide open source tools and protocols for image acquisition, registration, and visualization associated with SD-TIRF microscopy.
The setup is based on an inverted microscope connected to two scan units via independent ports - the left port is linked to the SD unit and the back port to scanner unit for TIRF and photo-activation/-bleaching experiments. Up to 6 lasers (405/445/488/515/561/640 nm) can be used for excitation. For excitation and detection of the fluorescence signal, either a 100x/NA1.45 oil or 60x/NA1.49 oil TIRF objective, respectively, have been employed. The emitted light is split by a dichroic mirror (561 nm long-pass or 514 nm long-pass) and filtered by various band-pass filters (55 nm wide centered at 525 nm, 54 nm wide centered at 609 nm for green and red fluorescence, respectively) placed in front of the two EM-CCD cameras. Please note that more technical details about the setup are listed in Zobiak et al.7. In TIRF configuration, the SD unit is moved out of the light path within circa 0.5 s so that the same two cameras can be used for detection, allowing faster switching between the two imaging modalities compared to circa 1 s that was reported in the past13. This feature enables dual-channel simultaneous acquisition, thus 4 channels SD-TIRF imaging at previously unmatched speed and accuracy can be performed. Moreover, alignment between SD and TIRF images is unnecessary. Image alignment between the two cameras, however, has to be checked before starting the experiment and corrected if necessary. In the following protocol, a registration correction routine was implemented in a self-written ImageJ macro. Moreover, the macro was mainly designed to allow a simultaneous visualization of SD- and TIRF datasets despite their different dimensionality. The acquisition software itself did not provide these features.