$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
Prerequisites for a valid and precise test result include correct, meticulous, and thorough pretest calibrations. For reports following a correct calibration with the vHIT system A, refer to Figure 1 and Figure 2. Calibration with the system B is done in one step for all six SCCs by asking participants to switch their gazes between the two dots that appear when the lasers are on (see Figure 3). Be careful to check that the eye and head velocities match after this calibration is done. A correct calibration includes a ∆ value below 21. For a detailed description of the calibration procedures, please refer to the manual provided by the manufacturer15,16.

Figure 1: Standard calibration prior to testing of horizontal SCCs with vHIT system A. It should be ensured that "Eye in Image" (image on the left) contains markings equivalent to the four outer limits as well as one in the center and that "Eye in Space" is depicted as a cross with vertical and horizontal lines in zero degrees. Please click here to view a larger version of this figure.

Figure 2: Head calibration for testing prior to testing of vertical SCCs with vHIT system A. Shown is a 3D representation of head movements with respect to the earth. Horizontal and vertical directions are shown together with head movements with respect to the possibly oblique axes of the inertial sensor. The three polar diagrams show the head movements from three different perspectives. Grey dots: raw head movement, black dots: calibrated head movement, solid grey line: camera orientation, solid black line: head orientation. Far left: grey and black dots must follow a line in the right-left direction (horizontal), middle: black and grey dots must follow a line in the superior-inferior direction (vertical), far right: black and grey dots must follow two perpendicular lines looking like a cross. Please click here to view a larger version of this figure.

Figure 3: Calibration procedure setup with vHIT system B. Ask the patient to position the left and right dots equidistant on each side of the fixation dot. As the procedure continues, only one dot at a time will be illuminated and the participant is asked to keep their gaze on the visible dot. As the participant’s gaze switches, the system tracks the movement of the pupil. Please click here to view a larger version of this figure.
Testing of the horizontal SCCs are done in a similar fashion with both types of equipment. For testing of the vertical SCCs, either the 2D or 3D test method may be used. Please refer to Figure 4 for a detailed description of the two test methods when testing all six SCCs.

Figure 4: Visualization of the vHIT test procedures. The left side illustrates the 3D vHIT procedure with vHIT system A. The right side illustrates the 2D modified vHIT procedure with vHIT system B. The middle section illustrates orientation of the semicircular canals (SCCs) being tested. The middle section illustrations are modifications of images taken from a smartphone application (see Table of Materials) and are used with permission from the copyright owner. For horizontal SCC testing, the examiner placed his hands on the patient’s jaw, delivering head impulses to each side. For vertical SCC testing, the examiner placed his dominant hand (in this study, both examiners were right-handed) on the top of the head and other hand beneath the chin. (a-c) Illustrations of the performance of the 3D vHIT using vHIT system A. In all three setups, the patient is facing the camera and the head is rotated in the direction of the SCCs being tested. (a) Right anterior left posterior (RALP) SCC testing. (b) Horizontal SCC testing. (c) Left anterior right posterior (LARP) SCC testing. (g–i) Starting position of the head; arrows illustrate the direction in which the head is rotated; the set of SCCs being tested is marked with red. (d–f) Illustrations of the performance of the 2D modified vHIT using vHIT system B. (d) RALP SCC testing with the subject turned 45° to the left and the impulses being delivered by either rotating the head forward or backward. (e) Horizontal SCC testing. (f) LARP SCC testing with the subject turned 45° to the right and the impulses being delivered by either rotating the head forward or backward. By rotating the patient’s head 45° prior to RALP and LARP testing, the eyes align with the axis of the vertical SCCs being tested; therefore, primarily vertical eye movements are produced when applying head impulses. (h), (j), and (k) show the starting positions of the head; the arrows illustrate the direction in which the head is rotated; the set of SCCs being tested is marked with gray. LARP indicates left-anterior-right-posterior plane; RALP indicates right-anterior-left-posterior plane. Reproduction of this figure has been granted with permission. Please click here to view a larger version of this figure.
Every time the vHIT test is performed, all individual steps of the test are important, as they may affect or alter test results. Following completion of every vHIT test, the examiner must go through the report meticulously to determine if results are valid. Special attention must be made to make sure that no noise or artifacts are included in the report. Eight different types of artifacts that may alter results have been described (see Figure 5). Even though the accompanying software removes a lot of noise and/or artifacts from the report, manual deletion of noise and/or artifacts may be needed as an additional step of the evaluation. If the test was carried out properly and participant cooperated fully during the testing, a conclusion of either a normal vestibular function or true compromised function may be drawn following evaluation and interpretation of the report. Please refer to Figure 6, Figure 7, and Figure 8 for test reports following examinations of participants with normal SCC function. Prerequisites for a normal complete vHIT test include mean gain values within the normal range as well as absence of pathological saccades. When mean gain values lie within the normal range, head and eye velocities are almost similar, and the corresponding curves are almost identical in the mirrored view. When no pathological saccades are present, the depiction of both head and eye velocities closely match both during and after application of the head impulses.

Figure 5: Visualization of eight different artifacts. Each type of artifact is illustrated by a graph as well as accompanying images depicting the test situation triggering the individual artifact (x-axis: time (seconds), y-axis: head and eye velocity (°/s)). Black and red lines indicate eye velocities and head velocities, respectively. The image on the left within a panel shows a subject being tested with vHIT system A, whereas the image on the right shows a subject being tested with vHIT system B. The appurtenant graph shows traces for eye and head movements related to the artifact. (a) Wrong calibration (high gain), (b) touching goggles (two peaks), (c) patient inattention (eye trace goes wrong direction), (d) bounce (head overshoot), (e) loose strap (delay/phase shift), (f) pupil tracking loss (trace oscillations), (g) mini-blink (pseudo-saccade), (h) blink (pseudo-saccade)17. This figure has been modified with permission17. Please click here to view a larger version of this figure.

Figure 6: Report with normal findings for lateral SCCs. It should be noted that the curves for both head and eye velocities match, all mean gain values are within the normal range (0.80-1.20), and there are no pathological saccades present. (A) vHIT system B report. Left: gain values are depicted as individual dots representing coherent pars of peak head velocities and gain values; red = right side, blue = left side. Mean gain values are also shown as a numerical value (0.91 and 1). Right: x-axis = time (milliseconds), y-axis = head and eye velocities. Head and eye velocities are shown in the same direction (mirrored view) to ease the interpretation. (B) vHIT system A report. Left: x-axis = time (milliseconds), y-axis = head and eye velocities (°/s). Head and eye velocities are shown in opposite directions. Right: gain values are depicted as a best fitted line through individual dots representing coherent pars of peak head velocities and peek eye velocities (first y-axis) as well as gain values (second y-axis); red = right side, blue = left side. Mean gain values are also shown as a numerical value (1.07 and 1.07). Please click here to view a larger version of this figure.

Figure 7: Report with normal findings for all six SCCs following vHIT system A testing. It should be noted that the curves for both head and eye velocities match, all mean gain values are within the normal range (0.80-1.20) or higher, and there are no pathological saccades present. Please click here to view a larger version of this figure.

Figure 8: Report with normal findings for all six SCCs following vHIT system B testing. It should be noted that the curves for both head and eye velocities match, all mean gain values are within the normal range (0.80-1.20), and there are no pathological saccades present. Please click here to view a larger version of this figure.
In order to conclude that the vestibular function is reduced a low mean gain value AND pathological saccades must be present. When low mean gain values are present, the amplitude of eye velocity is significantly lower than the corresponding amplitude of head velocity. Pathological saccades must also be present if the examination is truly pathological. These saccades might occur during or after the head movement. In order to conclude if saccades are truly pathological, the examiner must evaluate the saccades in terms of frequency, latency, direction, and amplitude. Please refer to Figure 9 and Figure 10 for examples.

Figure 9: Pathological test results after testing with the vHIT system A. Overt saccades are seen after head movement has stopped (A), covert saccades are seen during the head movement (B), and sometimes a mixture of both are seen (C). It should also be noted that mean gain values are below the normal range on the ipsilateral side of the pathological saccades. In (B) and (C), number 1 in red indicates covert saccades, number 2 in red indicates overt saccades, and number 3 in red indicate small correctional saccades that are classified as non-pathological saccades. Please click here to view a larger version of this figure.

Figure 10: Pathological test results after testing with the vHIT system B. Overt saccades are seen after head movement has stopped (A), covert saccades are seen during the head movement (B), and sometimes a mixture of both are seen (C). It should also be noted that mean gain values are below the normal range on the ipsilateral side of the pathological saccades. In (C), number 1 in blue indicates covert saccades and number 2 in blue indicates overt saccades. Please click here to view a larger version of this figure.