Teleoperated robotic system-assisted percutaneous transiliac-transsacral screw fixation is a feasible technique. Screw channels can be implemented with high accuracy owing to the excellent freedom of movement and stability of the robotic arms.
A subscription to JoVE is required to view this content. Sign in or start your free trial.
Method Article
Teleoperated robotic system-assisted percutaneous transiliac-transsacral screw fixation is a feasible technique. Screw channels can be implemented with high accuracy owing to the excellent freedom of movement and stability of the robotic arms.
Transiliac-transsacral screw fixation is challenging in clinical practice as the screws need to break through six layers of cortical bone. Transiliac-transsacral screws provide a longer lever arm to withstand the perpendicular vertical shear forces. However, the screw channel is so long that a minor discrepancy can lead to iatrogenic neurovascular injuries. The development of medical robots has improved the precision of surgery. The present protocol describes how to use a new teleoperated robotic system to execute transiliac-transacral screw fixation. The Robot was operated remotely to position the entry point and adjust the orientation of the sleeve. The screw positions were evaluated using postoperative computed tomography (CT). All the screws were safely implanted, as confirmed using intraoperative fluoroscopy. Postoperative CT confirmed that all the screws were in the cancellous bone. This system combines the doctor's initiative with the Robot's stability. The remote control of this procedure is possible. Robot-assisted surgery has a higher position-retention capacity compared with conventional methods. In contrast to active robotic systems, surgeons have full control over the operation. The robot system is fully compatible with operating room systems and does not require additional equipment.
The first robotic application utilized in orthopedic surgery was the ROBODOC system employed in 19921. Since then, robot-assisted surgical systems have rapidly developed. Robot-assisted surgery improves arthroplasty by enhancing the surgeon's ability to restore the alignment of the limb and the physiological kinematics of the joint2. In spinal surgery, the placement of pedicle screws using a robot is safe and accurate; it also reduces the surgeon's radiation exposure3. However, studies on robot-assisted surgery have been limited owing to the heterogeneity of traumatic orthopedic diseases. ....
Access restricted. Please log in or start a trial to view this content.
The application of this robotic technique was approved by the ethics committee of the Tongji Hospital of Tongji Medical College, Huazhong University of Science and Technology, and it complies with the Helsinki Declaration of 1975, as revised in 2013.
1. Preoperative planning
Access restricted. Please log in or start a trial to view this content.
A senior orthopedic surgeon completed the surgery using the procedure described. All the screws (three in S1 and two in S2) were secured. The time taken (from the first X-ray fluoroscopy to the insertion of the screw) for inserting each of the five screws was 32 min, 28 min, 26 min, 20 min, and 23 min, respectively. The fluoroscopy time for each screw was approximately 5 min. Although all the screws were in the correct place on the intraoperative fluoroscopic images, several articles have highlighted the need for postope.......
Access restricted. Please log in or start a trial to view this content.
Regardless of the type of Robot, the core application of robots in orthopedics provides an advanced tool for surgeons to improve the accuracy of surgery. However, the emergence of surgical robots is not a replacement for doctors. Surgeons performing robotic surgery may or may not be in the operating room. Surgical robots generally include a computer control system, a robotic arm responsible for the operation, and a navigation system responsible for tracking. There are three categories of robot systems depending on how th.......
Access restricted. Please log in or start a trial to view this content.
The authors declare that they have no competing interests.
| Name | Company | Catalog Number | Comments |
|---|---|---|---|
| 160-slice CT | United Imaging Healthcare Surgical Technology Co. Ltd | uCT780 | Acquire the prescise image and DICOM data |
| Electric bone drill | YUTONG Medical | None | Power system |
| Fluoroscopic plate base | None | None | Fix the cadaveric pelves to operating table |
| K-wire | None | 2.5mm | Guidewire |
| Master-Slave Orthopaedic Positioning and Guidance System | United Imaging Healthcare Surgical Technology Co. Ltd | None | A teleoperated robotic system that positions screws for orthopaedic surgery |
| Mimics Innovation Suite | Materialise | Mimics Medical 21 | Preoperative planning software |
| Mobile C-arm | United Imaging Healthcare Surgical Technology Co. Ltd | uMC560i | Low Dose CMOS Mobile C-arm |
| Operating table | KELING | DL·C-I | Fluoroscopic surgical table |
| Schanz pins | Tianjin ZhengTian Medical Instrument Co.,Ltd. | 5.0mm | Fix the cadaveric pelves |
| Semi-threaded screw | Tianjin ZhengTian Medical Instrument Co.,Ltd. | 7.3mm | Transiliac-Transsacral Screw |
| Seven DOF manipulator | KUKA, Germany | LBR Med 7 R800 | Device for performing surgical operations |