For demonstrative purposes, an assay designed to detect a commercially available, enhanced green fluorescent protein (eGFP)-expressing AAV2 vector, with a synthetic double-stranded DNA fragment containing eGFP as a quality control, was developed. Currently, there is ongoing debate as to whether the vector itself or a synthetic DNA fragment or linearized plasmid is most appropriate for use as the QC. Generally, a synthetic DNA fragment or linearized plasmid may be used if equivalency to the vector is demonstrated in method development (data not shown). Primers and probes were designed and optimized to detect the eGFP transgene. Refer to Supplementary Table S1 for the sequences used in this work. The concentration of the QC fragment stock was empirically determined using ddPCR. All assays were performed using the concentrations and PCR conditions given as examples in the protocol section.
For qPCR assays, it is recommended to evaluate the linearity, sensitivity, dynamic range, accuracy, and precision of the standard curve. Since ddPCR does not rely on a standard curve for target quantification, these recommendations must be modified. Instead, QCs consisting of synthetic double-stranded DNA fragments diluted to various concentrations to span the expected quantifiable range of a ddPCR reaction based on the Poisson statistical modeling were utilized29,30,31,32 to define the dynamic range and sensitivity and to evaluate accuracy and precision. The choice of QC concentrations was based primarily on the expected ratio of positive to total droplets within a well at a given concentration. Mathematically, ddPCR is theoretically most accurate when approximately 80% of the partitions positively amplify. As the positive to total droplet ratio increases above 0.8, the accuracy decreases due to saturation of the partitions, with quantification not being possible once 100% of the droplets are positive. On the low end, theoretically, as little as one positive droplet may be detected and quantified, though the accuracy is poorer and the assay is subject to low-level false positives. Typically, at least three droplets must be positive for a result to be calculated with 95% confidence, which is the threshold to calculate a concentration we used here.
A series of five different QC concentrations was prepared, with the target copy number/µL PCR reaction volumes calculated expected to yield positive to total droplet ratios spanning the quantifiable range of ddPCR, as shown in Table 5. These were used to evaluate the accuracy and precision of the assay. In the assessment here, the upper and lower limits of quantification were not pushed to the theoretical maximum possible in ddPCR. Accurate quantification may be possible at higher and lower levels than demonstrated here. The range should be developed in alignment with the downstream applications of this method.
A total of three independently prepared dilution series of these QCs were prepared in sample dilution buffer for each batch to evaluate the intra-assay accuracy and precision. Duplicate wells of each QC dilution were included. To simulate an actual validation protocol, a total of six accuracy and precision batches were performed by multiple analysts over multiple days. The results from these six batches were analyzed to define the intra-assay and inter-assay accuracy and precision of the method and to define the dynamic range of the assay.
Intra-assay performance was assessed for each batch at each QC level. We expected that all QC and NTC wells would have at least 10,000 droplets. This was met in 216 out of 216 wells tested across all six batches, with an average droplet count of 19,748 droplets/well (Table 6). Next, the inter-well %CV of each set of duplicate wells of each QC was expected to be ≤25.0%, except for the upper and lower limit QC, where ≤30.0% was expected. This was met in sets of 90 out of 90 wells tested across all six batches for the QCs, with an average inter-well %CV of 3.9% across all the QC levels (Table 7). All QCs yielded mean positive to total droplet ratios within the expected ranges outlined above (Table 6).
Within each batch, the intra-assay mean and standard deviation were calculated for each of the independently prepared dilution series points, and these were used to calculate an intra-assay mean for each concentration in each assay. This was used to assess the accuracy and precision of the assay (Table 8). Precision refers to the variability in the data from replicates of the same homogenous sample under normal assay conditions and is evaluated by calculating the %CV of the multiple included aliquots. We expected that the three aliquots tested within each batch would yield an intra-assay %CV ≤25.0%, except for the upper and lower limit QC where ≤30.0% was expected. This was met for all five QC levels in each of the 60 batches (30 out of 30 total performances). Generally, greater intra-assay precision than the target criteria could be achieved, with a mean intra-assay %CV of 7.7% across all QC levels. Accuracy refers to the closeness of agreement between the experimentally determined value and the nominal value. This is evaluated by calculating the percent relative error (%RE, or %Bias) between the calculated concentrations of each QC and their theoretically expected nominal concentrations. It was expected that the intra-assay mean of the three aliquots would be ±25.0% RE of the nominal concentration, except for the upper and lower limit QC where ±30.0% was expected. This was met for all five QC levels in each of the 60 batches (30 out of 30 total performances). Generally, greater intra-assay accuracy than our target could be achieved, with a mean absolute intra-assay %RE of 4.2% across all QC levels. In all performances of the NTC (30 total), no positive droplets were detectable.
Inter-assay accuracy and precision were also calculated using the intra-assay mean of each QC level within each batch. The inter-assay precision was expected to be ≤25.0% CV, except for the upper and lower limit QC where ≤30.0% was expected. Likewise, for inter-assay accuracy, ±25.0% RE was expected, except for the upper and lower limit QC where ±30.0% was expected. A significantly greater inter-assay accuracy and precision than these targets were observed (Table 9), with an inter-assay precision ranging from 4.0% to 8.5% and an inter-assay absolute accuracy ranging from 1.0% to 3.2%. Collectively, these results demonstrate that this method can achieve sufficient intra- and inter-assay accuracy and precision well within current industry targets. A dynamic range of this assay of 2,500-2.5 copies per µL of PCR reaction can be defined based on these results, with an overall assay sensitivity of 2.5 copies per µL of PCR reaction. As previously mentioned, it may be possible to validate wider dynamic ranges.
Next, it was necessary to evaluate assay accuracy and precision within the target matrix - in this case, tears. Typically, assays are validated prior to the initiation of clinical studies, meaning that tears collected from vector-treated patients are unlikely to be available for validation purposes. This can be artificially created by spiking the target AAV vector into tears collected from volunteer donors to create matrix-spiked QCs. Pooled human tears were collected by a third party (BioIVT). For proof of principle, an eGFP-expressing AAV2 vector acquired from a commercial source was utilized. The concentration of the AAV2 vector stock was empirically determined using ddPCR, without the use of a DNA isolation step, as described in this protocol. In each run, the AAV2 was independently spiked into the three tear aliquots at a high (expected 1.41 x 103 copies/µL PCR reaction) and low (28.2 copies/µL PCR reaction) level. Unspiked aliquots were included as a control to demonstrate the specificity of the method.
Intra-assay performance was assessed for each batch at each spike level. It was expected that all tear samples would have at least 10,000 droplets. This was met in 108 out of 108 wells tested across all six batches, with a mean total droplet number of 20,208 droplets/well (Table 10). Next, the inter-well %CV of each set of duplicate wells of each QC was expected to be ≤25.0% for the high and low spike levels. This was met in 36 out of 36 sets of wells tested across all six batches for the QCs, with a mean inter-well %CV of 3.2% (Table 11).
Within each batch, the intra-assay mean and standard deviation were calculated for each of the independently prepared tear spikes, and these were used to calculate an intra-assay mean for each concentration in each assay. This was used to assess the accuracy and precision of the assay in matrix (Table 12). We expected the intra-assay %CV to be ≤25.0% and high and low spike levels. This was met in six out of six batches for each level. Generally, greater intra-assay precision in matrix than the target could be achieved, with a mean intra-assay %CV of 3.7% at the high level and 12.2% at the low level (overall 8.0%). It was also expected the intra-assay %RE would be ±25.0% at both spike levels. This was met in six out of six batches for each level. Likewise, it was generally found that greater intra-assay accuracy in matrix than the target could be achieved, with a mean intra-assay absolute %RE of 8.1% at the low level and 11.3% at the high level (overall 9.7%). For the unspiked control, no eGFP signal was detectable in any of the aliquots (Table 12), demonstrating the specificity of the method in the human tear matrix.
Inter-assay accuracy and precision in tear matrix were also calculated using the intra-assay mean of each spike level within each batch. It was expected that the inter-assay precision would be ≤25.0% CV, and for inter-assay accuracy, we expected ±25.0% RE. A significantly greater inter-assay accuracy and precision than these targets were observed (Table 13), with an inter-assay precision of 5.5% at the high level and 7.1% at the low level, and with absolute inter-assay accuracy of 11.3% at the high level and 8.1% at the low level. Collectively, these results demonstrate the accuracy, precision, and specificity of the method in tear matrix.
Table 5: Quality controls used to define the dynamic range of the assay. Abbreviations: ULQC = upper limit quality control; HQC = high quality control; MQC = medium quality control; LQC = low quality control; LLQC = lower limit quality control; NTC = no template control. Please click here to download this Table.
Table 6: Total droplet counts and positive to total droplet ratios of synthetic double-stranded DNA quality control and NTC. Abbreviations: ULQC = upper limit quality control; HQC = high quality control; MQC = medium quality control; LQC = low quality control; LLQC = lower limit quality control; NTC = no template control. Please click here to download this Table.
Table 7: QC inter-well statistics (copy targets/µL PCR reaction). Abbreviations: ULQC = upper limit quality control; HQC = high quality control; MQC = medium quality control; LQC = low quality control; LLQC = lower limit quality control; NTC = no template control. Please click here to download this Table.
Table 8: Intra-assay accuracy and precision of QCs (copy targets/µL PCR reaction). Abbreviations: ULQC = upper limit quality control; HQC = high quality control; MQC = medium quality control; LQC = low quality control; LLQC = lower limit quality control; NTC = no template control. Please click here to download this Table.
Table 9: Inter-assay accuracy and precision of QCs (copy targets/µL PCR reaction). Abbreviations: ULQC = upper limit quality control; HQC = high quality control; MQC = medium quality control; LQC = low quality control; LLQC = lower limit quality control; NTC = no template control. Please click here to download this Table.
Table 10: Total droplet counts of tear samples. Please click here to download this Table.
Table 11: Tear sample inter-well statistics (copy targets/µL PCR reaction). Please click here to download this Table.
Table 12: Intra-assay accuracy and precision of tear samples (copy targets/µL PCR reaction). Please click here to download this Table.
Table 13: Inter-assay accuracy and precision of tear samples (copy targets/µL PCR reaction). Please click here to download this Table.
Supplementary Table S1: Sequences of primer, probes, and synthetic double-stranded DNA quality control utilized in this study. Please click here to download this Table.