This paper describes a striaghforward and efficient method of intubating mice for pulmonary function measurements or pulmonary instillation, that allows the mice to recover and be studied at later times. The procedure involves an inexpensive fiberoptic light source that directly illuminates the trachea.
A simple procedure to intubate mice for pulmonary function measurements would have several advantages in longitudinal studies with limited numbers or expensive animal. One of the reasons that this is not done more routinely is that it is relatively difficult, despite there being several published studies that describe ways to achieve it. In this paper we demonstrate a procedure that eliminates one of the major hurdles associated with this intubation, that of visualizing the trachea during the entire time of intubation. The approach uses a 0.5 mm fiberoptic light source that serves as an introducer to direct the intubation cannula into the mouse trachea. We show that it is possible to use this procedure to measure lung mechanics in individual mice over a time course of at least several weeks. The technique can be set up with relatively little expense and expertise, and it can be routinely accomplished with relatively little training. This should make it possible for any laboratory to routinely carry out this intubation, thereby allowing longitudinal studies in individual mice, thereby minimizing the number of mice needed and increasing the statistical power by using each mouse as its own control.
In 1999, Brown et al. published a paper describing a method for intubation of the mouse lung 1. Such a technique has considerable utility in doing repeat pulmonary function or bronchoalveolar lavage in individual mice in longitudinal studies 2. Since that original paper, there have been several other papers that have described different approaches to mouse intubation 3-9. While all of these methods can be used successfully, they usually require considerable training or cost. One of the main issues with such intubation is that as the intubation cannula approaches nears the trachea pending insertion, the cannula itself blocks the light and hence the visualization of where it needs to go. Thus, the insertion becomes blind at the most critical time. In this paper we show how to simply and inexpensively eliminate this visualization problem, thereby ensuring successful intubation with relatively little training or experience.
1. Preparing for the Procedure
One must first obtain and prepare the following items:
2. Performing the Intubation
As an evaluation of the method, we used four 20 week old male BALB/c mice with average weight (± SEM) of 27.7 ± 0.40 g. They were studied on five consecutive weeks, where the lung resistance was measured using a system as previously described 11. Each mouse was anesthetized with ketamine (100 μg/g BW) and xylazine (15 μg /g BW) in saline via IP injection. They were then intubated as described above. If there is any doubt whether the cannula is in the trachea and not the esophagus, this can be validated using a small dental mirror. Keep the mirror in a freezer, and when needed place in front of the Luer hub of the catheter. If the catheter is in the trachea, the exhaled breath will form a visible condensate on the mirror.
After intubation, we then connected the mice to the ventilator and measured Lung Resistance The mice were ventilated with a rate of 2 Hz and tidal volume of 0.2 ml, and respiratory resistance was measured by the inspiratory occlusion method as previously described11. Figure 5 shows 5 weekly measurements in each of the 4 mice. Reproducibility is excellent, showing that, at least at weekly intervals, there is no effect of the prior measurement. This is consistent with previously reported weekly assessments of mechanics and BAL cell profiles in individual BALB/c mice with a more difficult and potentially traumatic procedure 2.
Figure 1. Picture showing items used for intubation. The fiberoptic cable is shown inserted into a silicone rubber stopper, with a small piece of silicone rubber tubing tied near the opposite end. A silicone rubber stopper is attached to the light source as shown in Figure 2.
Figure 2. Picture showing rubber stopper connected to light source with other end of fiberoptic cable inserted through the intubation cannula. A simple support stand to hold the mouse during intubation is also shown on the left.
Figure 3. Two perspectives showing the position of mouse readied for intubation.
Figure 4. This figure shows an intubated mouse ready for ventilation. The tape around the mouth helps keep the cannula from moving. A small piece of Plasticine (modeling clay) provides a convenient rest to secure the cannula hub for connection to the ventilator.
Figure 5. Lung resistance from each of 4 mice (in different colors) measured at 5 weekly intervals.
Figure 6. Shown is a chart record of airway pressure in one intubated mouse following injections of 0.5, 0.75, and 1 ml. Each volume was held for 20-40 sec, then released before then next inflation. Although there may be a very slow leak after the stress relaxation recovery, this would have negligible effect on normal ventilation or assessment of dynamic pulmonary function measurements. Click here to view larger figure.
The procedure described here has several advantages. First the apparatus is simple and relatively inexpensive.. The fabrication of the apparatus does not require any special tools or costly equipment. The use of a catheter introduces that also is the light source means that one never loses sight of the tracheal opening as the introducer approaches the tracheal opening. The use of a 0.5 mm introducer also serves to minimize trauma that might occur with an initial insertion of a larger cannula. We note here that a similar optical probe is available from a commercial vendor (Braintree Scientific, Braintree, MA). Their device uses a battery powered light source and optical fiber.
In the present work, we tested the procedure with repeat measurement of lung mechanics, but such intubation could just as easily be used to instill chemicals or cells into the lung, as has been described for repeated delivery of LPS 12. In addition, a prior report with a more primitive intubation procedure described the ability to do repeated BAL in individual mice 2, and this would be much more simply accomplished with the new intubation approach.
In practice, the method described here has been easily taught to fellows, students, and technicians who had never attempted intubation. In fact, during group training sessions, some of the students become sufficiently proficient to then teach some of the other students who had not yet tried it. This method thus has a considerable advantage as it minimizes the number of mice needed for practice and should allow minimal damage in repeated studies.
In doing the intubation, there are several practical issues that should be mentioned. It is important to be as gentle as possible with the retraction of the tongue in the initial opening of the mouth. If unprotected forceps are used it is easy to injure the tongue, and this can lead to death of the mouse. In first learning how to do the intubation, the most important thing is the use of the finger behind the neck to adjust the angle of the head to enable visualization of the tracheal opening. When done correctly, with sufficient traction on the tongue, the vocal cords can easily be seen. It is this initial visualization step that usually requires the most time, since once the tracheal opening is seen, it is relatively simple to insert the fiber cable and intravenous catheter. Initially if there is an issue with this visualization, the trainee is often not pulling on the tongue with sufficient force. Increasing this pull slightly will straighten the visualization path so the vocal cords can be seen. Hamacher, et al., described a unique intubation system with microscopic visualization 4. Their online video of this intubation is excellent and very instructive, although the means of positioning the head and neck is not entirely clear from the video and figure. While the system they describe seems to work very efficiently, it requires a dedicated microscope. Using the system and procedure we describe, the vocal cords and tracheal opening can be seen with the naked eye. In our original description of this method 10, we described a procedure to add a cone to the intubation cannula. This cone wedges into the narrow mouse pharynx and prevents the cannula from being inserted too deeply. We have found that this wedge is useful in teaching students the procedure, since it is very easy to insert the cannula to the carina or beyond, possibly penetrating through an airway wall. Simple instructions to fabricate the wedge can be found in that paper. However, once someone learns the procedure sufficiently well and where to position the cannula, this adaptation is no longer needed.
Finally, we should note that we have only tested this procedure with the 20 g cannula in young adult mice of a few strains. In this situation, we have validated that the trachea and vocal cords can provide a very good seal around the cannula with normal ventilation pressures, i.e. there is minimal air leakage out of the lung with mechanical ventilation. Figure 6 shows results from on cannulation in a C57BL/6 mouse, where 3 increasing air boluses (0.5, 0.75, and 1 ml) were used to inflate the lungs. It is clear from this figure than pressure leaks are minimal to an airway pressure of at least 15 cmH2O. However, if one uses substantially younger or older mice, or mice from strains with a different lung anatomy, then it would be wise to confirm that there are minimal leaks. If there are, the procedure may then require use of a different size cannula.
In summary, the intubation procedure describe here is inexpensive to fabricate and simple to use, and it should enable most investigators and laboratory technicians to quickly learn to successfully intubate mice with relatively little experience.
The authors have nothing to disclose.
Supported by NIH HL-10342.
Material | Company | Catalog # | Comment |
Intubation cannula | BD Insylte, Sparks, MD or Jelco Optiva, Carlsbad, CA | 1-in.-long, 20-gauge intravenous (IV) catheter | |
Fiber-optic cable | Edmund Optics, Barrington, NJ | #NT02-542 | Approximately 2-ft length of 0.5-mm optical fiber (Communication grade plastic fiber). The edge of the fiber end that is inserted into the trachea should be gently rounded by holding the fiber ≈2 cm from the end and then making small circles while dragging the tip for a few seconds on 1,000-grit emory paper. |
Light Source | Volpi | NCL-150 | Although we use a 150-W halogen light source, any equivalent light source, even with much lower wattage can be used. |
Silicone rubber stopper | A #1 silicone rubber stopper fits the light source used in this demonstration. Different light sources may require a different size. | ||
Small silicone rubber tube | Cole-Palmer | EW-96410-13 | A ≈1.5 cm piece of silicone rubber tubing (0.8 inner diameter, 4 mm outer diameter) |
Angled support stand | Ours is constucted from plexiglass, but any material to which a thread or wire can be affixed to hold the mouse at an almost vertical angle can be used. |