The following protocol describes the major steps of the laser ablation electrospray ionization (LAESI) experiment and provides representative examples for lateral and three-dimensional (3D) imaging for animal and plant tissue samples. Additional experimental and technical details can be obtained from elsewhere.1-6
1. Tissue Preparation and Mounting
- If sectioning is needed, use a cryomicrotome to section the tissue into 10-100 μm thick slices at -10 to -20 °C unless otherwise recommended for a particular tissue type.
- Mount sections onto a flat surface (e.g., chemically pre-cleaned glass slide) directly without chemical modifiers. For sectioned tissues, thaw-mount the sections and secure the sample holder to a Peltier-cooling stage immediately upon thaw-mounting to keep the tissue frozen at all times during analysis. This step is needed to minimize/prevent molecular migration in the section.
- If needed, use a heat sink equipped with a low-power fan to facilitate heat removal from the Peltier-stage to maintain the tissue frozen.
- In a humid environment over an extended period of time (1-2 hours), inspect for condensation of water or ice on the tissue surface. Water condensation on the tissue detrimentally affects the imaging performance in LAESI experiments.1
- If needed, use a room dehumidifier or place the cooled sample in an environmental chamber filled with an inert gas (e.g., dry nitrogen gas) to prevent condensation.1
2. Optimization of the LAESI Ion Source
The LAESI ion source consists of a mid-infrared laser, a series of optical elements for light steering and focusing as well as additional sample holders, cooling components, translation stages, and a solvent delivery system. Figure 1 demonstrates the typical arrangement of these elements with respect to the entrance of the atmospheric-pressure ion source of a mass spectrometer.
- As illustrated in Figure 1, position the sample 15-20 mm below the orifice of the mass spectrometer sampling cone (dOR-FP).
- Operate a mid-IR laser at 2.94 μm wavelength and 10 Hz repetition rate. Attenuate the laser output to ~100 μJ/pulse energy.
- Use a combination of gold mirrors and a focusing lens transparent at the laser wavelength (e.g., a plano-convex CaF2 or ZnSe lens) to couple the laser pulse energy into the sample at normal incidence (see right incidence angle with respect to sample surface in Figure 1).
- Position the mid-infrared beam axis 5-8 mm in front of the orifice of the mass spectrometer sampling cone.
- Adjust the focusing lens position and the pulse energy of the laser beam to achieve tissue removal in the focal spot. The dimensions of the ablated volume determine the pixel (or voxel in three-dimensional imaging) size for the imaging application.
- Position a nanospray emitter in line with the inlet axis of the mass spectrometer and at an orifice-to-emitter tip distance of ~10 mm (see Figure 1).
- For the electrospray, prepare 50% methanol solution with 0.1% acetic acid or 0.1% ammonium acetate additive for positive or negative ion mode, respectively. Depending on the sample, other organic solvents, such as acetonitrile, isopropanol, etc., may replace methanol at concentrations appropriate for the analytical task. The stability of the electrospray is crucial for successful imaging. Depending on the solvent selection, the flow rate and the spraying voltage need to be adjusted to achieve stable spray.
- For reactive LAESI6 in imaging applications, the electrosprayed solution may contain reactants.
- Use a syringe pump to deliver the electrospray solution through the electrospray emitter at a flow rate of ~300 nL/min.
- If the mass spectrometer orifice is kept at a low voltage (<500 V measured against ground), generate electrospray by applying high voltage directly to the electrospray emitter (e.g., 3000 V) or through a metal union. Otherwise, ground directly the electrospray emitter or through metal union to establish electrospray.
- Operate the electrospray source in cone-jet spraying mode for most efficient ion generation by LAESI. For the effect of operating variables on the spraying modes and their effect on the mass spectra, see discussions elsewhere.5,7,8
- Carefully adjust the relative distances of the LAESI setup to optimize for LAESI ion yield while keeping the laser beam, the emitter, and the orifice axes in the same plane. Please find detailed instructions elsewhere.5
- With an optical microscope, determine the lateral dimensions of the ablation crater on the sample.
- For three-dimensional LAESI imaging experiments, perform ablation with individual pulses and determine the depth of a voxel using, for example, the z-stack mode in optical microscopy.3
3. Molecular Imaging and Data Analysis
In the imaging experiment, the tissue sample is moved in the focal plane of the laser in X and Y directions with step sizes larger than or equal to the dimensions of the ablation spot. The spatial resolution is limited by the focusing of the incident laser beam.
- Select the area of interest on the sample surface and obtain the (X, Y) coordinates of the corresponding boundaries.
- Chose a gridding algorithm (e.g., adaptive grid, selected region imaging, rectangular grid, spiral pattern, Z scanning, etc.) with which to raster the sample surface with selected dwell time at each pixel over the area to be imaged.
- Use a three-axis translation stage and software that is capable of rastering the sample according to the predetermined grid.
- Calculate the total time required for imaging.
- Disable the data acquisition time limit of the mass spectrometer. If this is not possible, set the data acquisition time limit to the calculated imaging time value.
- Start the mid-infrared laser source at a repetition rate proper to produce sufficient signal-to-noise ratio in the mass spectrum within the dwell time at each pixel to perform a LAESI lateral imaging experiment. For 3D molecular imaging, use a spectrum acquisition rate higher than the laser source repetition rate to successfully mass-analyze the ions generated within a single laser pulse. Wait for the START signal to initiate the ablation sequence.
- Turn on the electrospray source. Make sure that there is enough solution for the full time required for imaging.
- Simultaneously START the acquisition of mass spectra, the mid-IR laser ablation, and the surface scanning.
- When the imaging run has finished, STOP the surface scanning, the mid-IR lasing, and the data acquisition.
- Disable the laser source.
- Turn off the high voltage.
- Switch off the syringe pump.
- Set the mass spectrometer to STANDBY mode.
- Turn off the Peltier-cooling electronics.
- Close inert gas flow if used.
- Use a software to correlate the absolute coordinates of the pixels in lateral imaging or the voxels in 3D analysis with the corresponding spectra.
- Plot the ion intensity signal for a selected m/z value against the absolute coordinates of analysis to obtain lateral and 3D molecular images.
4. Representative Results
Figure 2 gives representative results for some major tissue types and imaging modalities. Panel A depicts a case for an animal tissue section that has been frozen during the experiment to prevent dehydration.1 In addition, the sample was located in a dry nitrogen gas environment to avoid ambient water vapors from condensing on the sample surface. A 100-μm-thick coronal section of a rat brain (Rattus norvegicus) was laterally imaged with LAESI. The anatomical regions of the brain (see optical image in panel A) show good correlation with the molecular image obtained for the plasmalogens PC(O-33:3) and/or PE(O-36:3) with m/z 728.559.
Panel B shows the 3D LAESI imaging of a Zebra plant (Aphelandra squarrosa) leaf tissue. Because leaves possess a natural defense mechanism against dehydration, the sample could be interrogated in the ambient environment.3 The obtained 3D molecular images revealed a variety of distribution patterns for primary and secondary metabolites. Among others, acacetin with m/z 285.076 was detected at higher ion counts in the yellow sectors of the second and third layers from the top with a homogeneous distribution in the others. This distribution agreed with the pattern of the variegation seen in the optical image.

Figure 1. Schematic of the LAESI system (ES, electrospray emitter tip; OR, orifice of mass spectrometer sampling cone; FL, focusing lens; FP, focal point; P, Peltier cooling stage; HS, heat sink). A portion of the particulate matter expelled during mid-IR ablation (red dots) coalesces with the electrospray to yield charged droplets seeded with molecules and ions of the sample (green dots). The ions released from these droplets are analyzed and recorded by the mass spectrometer.

Figure 2. Representative results for lateral and 3D imaging with LAESI mass spectrometry. (A) The top panel depicts the optical image of a rat brain (Rattus norvegicus) coronal section and the molecular image obtained for the plasmalogens PC(O-33:3) and/or PE(O-36:3) with m/z 728.559. The white scale bar corresponds to 1 mm. Adapted with permission from (Reference 1). Copyright 2010 American Chemical Society. (B) The bottom panel shows the 3D imaging of a leaf from a variegated Zebra plant (Aphelandra squarrosa). Acacetin with m/z 285.076 was detected at higher ion counts in the yellow sectors of the second and third layers from the top with a homogeneous distribution in the others. Reprinted with permission from (Reference 3). Copyright 2009 American Chemical Society.