$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
Dextran
A successful coating of dextran should produce a fluorescent monolayer. At the high concentration this should appear relatively uniform. At lower concentrations it will appear sparser (Figure 2A). Many STORM/PALM microscopes have the ability to change the angle of illumination from epifluorescence to TIRF. When using a full frame camera view, for example at 512 x 512 pixels on a typical EMCCD camera, an even illumination should be observed. If there are any stripes or out-of-focus regions this can indicate that the sample should be reinserted into the sample holder with fresh oil on the objective lens. Alternatively it may indicate a problem with the microscope.
When acquiring super-resolution raw data the imaging laser should be increased in power to approximately 2 kW/cm2. There will be an initial burst of fluorescence followed by blinking as the fluorophores are driven into temporary dark states. At high dextran densities the blinks are likely to overlap, particularly near the beginning of the image acquisition (Video 1). At medium and low densities these blinks should be sparse, i.e. spatially separated, and in focus with no obvious background (see frames 2,000, 5,000 and 8,000, Figure 2A, Videos 2 and 3). During this image acquisition phase, at constant laser illumination, the number of blinks will decrease with time (compare frame 2,000 with frame 8,000 in the high density sample, Figure 2A). The main difference between the various super-resolution images of the different dextran concentrations (high, medium and low) is the decreasing number of localizations (Figures 2A and 2B). In other words, the concentration of the fluorescent molecules, number of blinks in the raw data and the number of localizations have a proportional relationship. This relationship, however, is not a simple linear one as at very high molecule densities the software is unable to successfully localize molecules (compare the red and blue lines, Figure 2C). The reason for this is that at the high concentration it is not possible for the processing software to fit the positions using the Gaussian fitting algorithm where the blinks are non-sparse. As the blinks decrease through the acquisition phase due to photobleaching the software can fit more and more of the blinks as they become sparse (Figures 2A and 2C). Moreover, individual dye molecules may blink, and therefore be localized more than once 28,41,42 .
A common problem when imaging any of these samples, but particularly the high density dextran one, can be the presence of bright but unfocused fluorescent haze that appears to be rapidly diffusing during the STORM image acquisition phase. This is distinct from the high contrast fluorophores on the surface of the glass, which can be seen blinking (Figure 3A, Video 5). These detached fluorescent molecules can be prevented or removed by increasing the number of wash steps with PBS prior to adding the switching buffer or by adding fresh switching buffer to the chamber (Figure 3B, Video 6). Processing and comparing the data from each image sequence results in very different super-resolution images which have a decrease in both the number of localizations and the precision with which they can be fitted of the software to localize the blinks (Figures 3C, 3D, 3E and 3F).
Actin Filaments
Preformed actin filaments can be seen stuck to the surface of the glass by diffraction-limited imaging (Figure 4A-4C). If the number of filaments appears very low, then a longer incubation time can be used or a decrease in the volume of the actin filament solution also helps. Selecting single relatively bright filaments (Figure 4D) rather than tangled areas results in better quality images. During the acquisition phase, bright in-focus blinks should be seen along the length of the filament (Video 7). Sparse blinking should be seen during the acquisition phase and then subsequently in the processed data there should be a thin continuous filament (Figure 4E) and the localizations per frame should a gradual decline (Figure 4F), unlike in Figure 3E.
If blinking is non-sparse, or if the software not being able to localize the molecules, a more subtle artifact, called a mislocalization 32, can result. This arises when the software averages the position of two overlapping molecules and the localization is position mid-way between them. An indication that there have not been a significant number of overlapping blinks, and consequently mislocalizations, is that the mean precision estimates should be the same in the row and column directions, i.e. horizontally and vertically; where there are mislocalizations these would have occurred along the length of the filament, produced a larger than normal blink (i.e. spread over more pixels), which would consequently have been localized with less accuracy in one of the directions. This is most easily seen where there is a single actin filament in the imaging area and it is lying horizontally or vertically. In this case, the STORM microscope, achieved a mean precision of 16 nm in both directions. This results in a precision limit, a measure of effective image resolution of approximately 34 nm. These metrics are provided by rainSTORM 27, however, as we have a filamentous structure of uniform diameter, 7 nm with the very small phalloidin-Alexa 647 label we can take a measure of FWHM to estimate the resolution of the image. By drawing a straight line through the filament of the super-resolution image in ImageJ (Figure 4G), using the plot profile feature (Figure 4H) and subsequently performing a Gaussian fit (Figure 4I) the FWHM is calculated as 43.2 nm. When attempting this measurement we recommend that the pixel size should match or be slightly smaller than the mean precision estimate for the image (see info.txt file Figure 1G). In this case, 16 nm pixels were used to reconstruct an image.
Two related problems can occur with STORM, where the blinking fluorophores become non-sparse, i.e. individual molecules within approximately 250 nm blink at the same time and therefore the signals overlap. The first is that depending on the algorithm and the quality control criteria it uses, the blinks may not be localized at all. This results in super-resolution images with few or no localizations. The second problem of mislocalizations occurs when the two blinks occurred sufficiently close to each other to appear as a single blink. In this case the position in the final image represents an average of the two. For more detail on this see reference 32. In both cases this can occur with very high density samples, insufficient laser power or with switching buffer problems. This problem is apparent where a number of actin filaments are branching and/or crossing each other (Figures 5A-D, Video 9). By processing subsets of frames, and comparing the first and last 5,000 frames we can see a different resulting image. In the super-resolution image using the first 5,000 frames (Figure 5C) we can see that there are many localizations in the middle of the image, however when using the last 5,000 frames, very few of these are apparent and we are left with just the filaments, albeit somewhat discontinuous owing the low number of localizations in the image (Figure 5D). If a too high blinking density is suspected comparison of the images with the localization per frame data can strongly suggest that this problem is occurring; during the first set of frames there is an average number of localizations per frame of over 10 compared with the last set of frames where it is approximately 4 (Figure 5E). In order to minimize the likelihood of mislocalizations occurring it is ideal to have as few localizations per frame as possible, although if there are a large number of molecules to be imaged, i.e. for a high density sample, this requires that a large number of acquisition frames be taken leading to another problem in STORM microscopy which is drift.
Lateral Drift - Actin Filaments and Fiducial Markers
Drift occurs when the sample moves in relation to the objective lens through the data acquisition phase. This is difficult or impossible to see during the image acquisition phase, particularly if it is lateral rather than axial, or unless it is being specifically measured, as it is typically less than 50 nm over the course of several minutes for relatively stable microscope systems. However, in reconstructed images of known structures such as actin filaments with well-defined structure, it can be detected in the super-resolution images. The first sign that lateral drift may have occurred is that the structure is larger than expected (Figure 6A, Video 8) for example with a relatively large FWHM compared with the precision limit data from rainSTORM, in this case 90 nm compared with 67 nm. However, a better way to detect drift is by comparing the localizations as a function of time, i.e. seeing if the localizations in the later frames are displaced compared with those in early frames. This can be clearly seen in the case of actin filaments, which are very small and of uniform structure, when displayed with a color code using the box tracking feature in rainSTORM (Figures 6B and 6C).
Drift is a well-recognized problem and there are a number of strategies that can be used to minimize it 19 or correct it post-acquisition either using fiducial markers 21,43 or cross-correlations 44 . In order to measure drift and correct for it, fluorescent beads of 100 nm diameter can be used as fiducial markers (Figure 6D). Because they are small and bright their position can be accurately measure using Gaussian fitting algorithms, such as rainSTORM. In an example where there is relatively severe drift of approximately 100 nm over the course of 3 min 10 sec, during the acquisition phase (Figure 6E), the same box tracking feature can be used to color code and confirm that drift occurred (Figure 6F). As all four beads in this example show near-identical drift it is possible to select one of them, in this case bead 2, to use as reference and subtract the drift from the other beads (Figure 6G). By adding fiducial markers to a biological sample of interest it then becomes possible to measure and correct any drift where the underlying structure is unknown or far more variable than an actin filament or a fluorescent bead.
Epidermal Growth Factor
Finally, EGF-stained HeLa cells can be used to give a realistic example of image resolution in cells (Figure 7A, Video 10). These cells are relatively straightforward to image as they should have the majority of the EGF-fluorescence in the plane-of-focus on the cell surface in close proximity to the coverglass. The less bright region the in center corresponds to the position of the nucleus. TIRF illumination can enhance the image quality by eliminating out-of-focus fluorescence coming from parts of the cell membrane not in close proximity to the glass (approximately 150 nm penetration into cells). Zoomed in regions of interest in the diffraction-limited image are typically indistinct (Figures 7B and 7C), however in the super-resolution images there should be a mixture of clusters and occasional isolated single pixels (Figures 7D-7F). These will represent either isolated EGF receptors at the cell surface or possible a small amount of non-specific binding. The clusters will be approximately 100 nm in diameter and are likely to correspond to forming pits and vesicles, the pathway via which EGF is predominantly down-regulated and endocytosed. Typical mean precision estimates are around 20 nm for this type of sample with a precision limit of approximately 45 nm. It should be noted that this precision limit measure of effective resolution does not take into account label size, or any drift, which can be measured with fiducial markers, but is still noticeable by "comet tails" on the clusters or using the box tracking feature as outlined in Figure 6 and described in protocol section 8.
| Component | Final Concentration |
| Catalase | 1 μg/ml (50 units) |
| Glucose | 40 mg/ml |
| Glucose oxidase | 50 μg/ml |
| Glycerin | 12.50% |
| KCl | 1.25 mM |
| MEA-HCl | 100 mM |
| TCEP | 200 μM |
| Tris | 1 mM |
Table 3. Switching buffer
| Typical Acquisition Settings | Value |
| Pixel size (nm) | 160 |
| Exposure time (msec) | 10 |
| Gain | 200 |
| Frame size (pixels) | 128 x 128 |
| Cycle time (fps) | 52.5 |
| Frame number | 10,000 |
| 640 nm laser power density (kW/cm2) | 2 |
Table 4. Acquisition settings

Figure 1. STORM Image Reconstruction Using rainstorm. (A) Opening rainSTORM from within MATLAB. (B) rainSTORM graphical user interface (GUI). (C) rainSTORM Reviewer GUI. (D) Adjust contrast window. (E) STORM image after review and contrast adjustment. (F) Histograms of image quality metrics. (G) Files generated after pressing save image button in reviewer GUI. Click here to view larger figure.

Figure 2. (A) Diffraction limited (DL) images show different concentrations of dextran prior to STORM imaging. Super-resolution (SR) image reconstructions have a pixel size of 25 nm. 10,000 images were collected with a 128 x 128 pixel frame size and individual frames are shown from that sequence (2,000, 5,000, 8,000). Acquired with a 10 msec exposure time at 52.5 frames per second. Images have a pixel size of 160 nm. For clarity of viewing a 0.1% pixel saturation contrast enhancement was applied using ImageJ. The mean precision estimates are 28 nm (high), 24 nm (medium) and 16 nm (low) for the different dextran images. The localization densities are 724 per μm2 (high), 526 per μm2 (medium) and 13 per um2 (low). (B) Graph showing an average of three 10,000 frame sequences of each concentration of dextran. Error bars show standard deviation.(C) Graph plotting the number of accepted localizations per frame using a rolling 100 frame average. Click here to view larger figure.

Figure 3. Poor Quality Dextran Data. (A) Diffraction limited frame from a 15,000 frame STORM sequence. Note the diffuse fluorescent haze across the image. This is caused by fluorophores diffusing through the medium. 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. (B) The same as (A) after the buffer had been changed. Note the improved contrast as unbound fluorophores have been washed away. (C) A super-resolution image reconstruction corresponding to the data collected in (A). Identical image size but with pixels of 25 nm. The mean precision estimate is 35 nm. (D) A super-resolution image reconstruction corresponding to the data collected in (B). Identical image size but with pixels of 25 nm. The mean precision estimate is 30 nm. (E) Graph plotting the number of accepted localizations per frame, using a rolling 100 frame average. The red line corresponds to STORM sequence (A & C) where there is a high background. The blue line shows data corresponding to (B & D), where the background is low. (F) Graph showing the accepted localization number for the images corresponding to high (C) and low (D) background data. Click here to view larger figure.

Figure 4. Typical Actin Data. (A-C) Diffraction-limited images of actin filaments prior to addition of switching buffer and STORM imaging. Variable lengths of filaments can be seen. Very bright filaments are often several filaments tangled together. Pixel size is 160 nm. (D) A diffraction-limited image of a single actin filament. (E) STORM image (0.1% contrast enhancement applied in ImageJ). From a 10,000 frame sequence with a 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. The pixel size is 16 nm. The mean precision estimate is 16 nm. (F) Graph plotting the number of accepted localizations per frame, using a rolling 100 frame average. (G) A zoomed in region of the actin filament from (E) prior to contrast enhancement with a yellow line profile drawn across. (H) A plot profile view from the yellow line drawn across the actin filament. Generated in ImageJ. (I) A Gaussian fit of the plot profile using the curve fitting tool in ImageJ. The standard deviation, d, was multiplied by 2.35 to get a FWHM measurement of 43.2 nm. Click here to view larger figure.

Figure 5. Mislocalized Actin Filaments. (A) Diffraction-limited actin filament. 160 nm pixels. (B) STORM image of actin filament shown in (A). From a 20,000 frame sequence with a 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. The STORM pixel size is 16 nm. All 20,000 frames were processed. The mean precision estimate is 17 nm. (C) As (B) but with only the first 5000 frames of the 20,000 frame sequence processed. The mean precision estimate is 18 nm.(D) As (B) but with only the last 5,000 frames of the 20,000 frame sequence processed. The mean precision estimate is 17 nm.(E) Graph of localizations per frame data from full 128 x 128 pixel frame view.

Figure 6. Lateral Drift. (A) STORM image of an actin filament reconstructed from a 10,000 frame sequence with a 64 x 64 pixel frame size, a 10 msec exposure time and taken at 64.6 frames per second. The STORM pixel size is 32 nm. The mean precision estimate is 32 nm. (B) A STORM image displayed using the 'Box Tracking' feature in rainSTORM. Localizations are displayed with a color corresponding to the frame number of when they were acquired, for example, localizations from early in the acquisition sequence are blue; localizations from late in the acquisition sequence are red. The displaced colors indicate that drift has occurred. (C) A zoomed in region of (A) displayed using the Box Tracking feature in rainSTORM. The displaced colors indicate drift has occurred. (D) A diffraction limited image of four 100 nm fluorescent beads, which can be used as fiducial markers. Pixel size 160 nm. Numbers 1-4 show cropped and zoomed beads individually. (E) Each fluorescent bead corresponding to (D) reconstructed in rainSTORM from a 10,000 frame sequence with a 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. The STORM pixel size is 5 nm. The mean precision estimate is 7 nm. (F) Beads corresponding to (D) and (E), displayed using the 'Box Tracking' feature. The displaced colors indicate drift has occurred. (G) Using bead number 2 as the reference, beads 1, 3 and 4 are displayed after the 'Subtract Drift' feature is used in rainSTORM. Comparison with (E) shows that the lateral drift has been corrected. The STORM pixel size is 5 nm. Click here to view larger figure.

Figure 7. Typical EGF Data. (A) Diffraction limited image of part of a HeLa cell focused at the basal cell surface. Yellow boxes indicate zoomed in regions of interest shown in (B) & (C). (B) Zoomed in diffraction limited image from the region near the edge of the cell (box B). (C) Zoomed in diffraction limited image from region under the nucleus (box C). (D) STORM image corresponding to (A). From a 10,000 frame sequence with a 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. The STORM pixel size is 25 nm. The mean precision estimate is 21 nm. (E) STORM image corresponding to box E (D). (F) STORM image corresponding to box F (D). (G) Localizations per frame data from (D). Click here to view larger figure.
Videos 1-4. Videos correspond to Figure 2A with varying dextran concentrations: 1= high, 2= medium, 3= low, 4 =none). 10,000 frame sequences with 128 x 128 pixel frame sizes, acquired with a 10 msec exposure time at 52.5 frames per second. Click here to view video 1, video 2, video 3, video 4
Videos 5-6. Videos correspond to Figure 3 A & B. Video 5 is pre-wash and Video 6 is post-wash after unbound fluorophores have been removed. 15,000 frame sequences using a 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. Click here to view video 5, video 6.
Video 7. Video showing the raw frame sequence that was processed to generate the STORM image in Figure 4E. 10,000 frame sequence with a 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. Click here to view video 7.
Video 8. Video showing the raw frame sequence that was processed to generate the STORM image in Figure 5B. 20,000 frame sequence with a 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. Click here to view video 8.
Video 9. Video showing the raw frame sequence that was processed to generate the STORM image in Figure 6A. 10,000 frame sequence with a 64 x 64 pixel frame size, a 10 msec exposure time and taken at 64.6 frames per second. Click here to view video 9.
Video 10. Video showing the raw frame sequence that was processed to generate the STORM image in Figure 7D. 10,000 frame sequence with a 128 x 128 pixel frame size, a 10 msec exposure time and acquired at 52.5 frames per second. Click here to view video 10.