SYSTEM MONITORING
What things look like when experiment is running well:
Once the cannulae have been placed in the circuit and the lungs are ventilating, there are multiple ways to ensure the system is working properly. There should be no leaks of perfusate throughout the line. The pulmonary vascular resistance (PVR) should remain relatively constant (assuming a constant flow). The oxygen exchange should increase once the ventilator is working properly and expanding the lungs to recruit more alveoli for gas exchange. Figure 12A shows properly ventilated and perfused lungs connected to the EVLP circuit inside the artificial thorax.
What things look like when experiment is not running well:
There are a few common issues that have had the highest rate of occurrence during the beginning stages of an EVLP experiment. The first and easiest to remedy is a leak in the line exiting from the lung. This is noticeable by a pool of perfusate pooling under part of the circuit and the level in the reservoir continually decreasing. Check and tighten any tube connectors around the spill area and inspect the tube itself for a leak. If this leak occurs before the lung, it may also introduce bubbles into the lung. This should be remedied as quickly as possible as air bubbles in the perfusate will result in tissue damage and cause a significant increase in the PVR. There may also be a leak coming from the lung or one of the cannulae. This may be caused by either slippage of a cannula or an obstruction in the exiting line causing a pressure buildup. Inspect the position for both cannulae to ensure neither has slipped or twisted. The PA pressure should also be monitored during this process because an instantaneous increase in PA pressure is an obvious sign that an obstruction of some kind has recently occurred. Figure 12B shows a ruptured lung that ruptured due to high pressures. A leak from the lung itself may also be caused by a tear in the tissue. This issue may or may not be reparable but repositioning and retightening the cannulae is the best option in this scenario.
Key Learning Points/Opportunities:
Trial and error development of the ex-vivo lung perfusion system has allowed us to identify several key issues that we outline here to facilitate efficient implementation of the EVLP system. First, with respect to procurement, it is important that standard anesthetic techniques are followed to properly anesthetize the animals (enough anesthetic, injection into the peritoneum) and adherence to all IACUC policies is required. The cannulae (shown in Figure 13 A, B, and C) should be repetitively flushed in order to remove any clot and/or debris within the pulmonary vasculature. With respect to animal selection, we suggest using Sprague Dawley or Lewis rats weighing 250–350 g. Special care should be taken when cannulating rats weighing close to 250 g since the vessels will be smaller and therefore much more difficult to cannulate without injuring the vasculature. If smaller rats, or a mouse model, is to be used, smaller cannula may need to be used.
Tracheal cannulation is not typically challenging as long as the suture is secured properly by first passing a silk suture posterior to the trachea after dissecting the surrounding fascia and prior to cannulation. Follow this with an anterior incision 1-2 tracheal rings above the suture to pass the cannula. Tie square knots in between the tracheal rings in order to secure it within a groove for better security (Figure 4C). Cannulation of the pulmonary artery (PA) is more challenging compared with the tracheal cannula. The following steps were used in this study for this procedure. First, grasp the heart apex with a pair of forceps. Pass another pair of forceps in the transverse sinus and thread a suture to secure the cannula in the proximal PA. Incise the right ventricle immediately before the right ventricular outflow tract (RVOT) (Figure 14A). After the incision into the RVOT, the cannula will be guided towards the pulmonary artery outflow tract. Having the suture in position behind the pulmonary artery/aorta before the right ventriculotomy increases efficiency (Figure 5C). The cannula should be secured into place with the suture to prevent dislodgement. A major complication can occur if the PA cannula is not in the correct anatomical orientation. The cannula may be inserted too far and only perfuse one branch or become mal-positioned with twisting of the heart-lung specimen upon removal from the chest cavity. This can easily be oriented back to the original position to preserve the proper angle of anatomical position. Finally, Left Atrial (LA) cannulation is the most challenging part of the procedure. The LA cannula needs to be placed within the left atrium. With the tissues being extremely friable, be mindful to not use significant force or twisting in order to prevent a tear within the pulmonary vein & left atrium which would then make the experiment unsalvageable. The PA cannula is best placed before the LA cannula. A left ventriculotomy with removal of the apex has been shown to disrupt the cordae tendinae and allow easier access through the mitral leaflets. Also, the ventriculotomy makes it easier to dilate and visualize the mitral valve and to feed the cannula through the mitral valve. Dilation of the mitral valve annulus with a pair of small blunt-ended pick-ups can be done in order to visualize the tract into the LA (Figure 14B). Suture should be placed behind the heart before cannulation. This can be done simply by lifting the heart up using a pair of small blunt-ended pick-ups and placing the suture underneath and across the heart. The LA is now ready to be cannulated. Feed the LA cannula through the pick-ups in order to properly visualize the placement of the cannula into the left atrium. Take special care not to dislodge the cannula back into the left ventricle. The suture should then be tightly secured along the myocardium of the left ventricle. Securing the suture to the left atrium could occlude the entire or part of the cannula.
During the procedure, it is crucial that no air remain in the inflow section of the apparatus. Any significant air can produce an air embolism increase the PVR (effectively an “air-lock”) which will result in a much lower perfusate flow for a given pressure. Various points can be used to remove air within the system. Air within the outflow section is expected and should not have any deleterious effect on the lungs. A porcine model for pulmonary hypertension has been shown to recreate the pathology from continuous small amounts of air over an 8-week period. The increased air decreases the amount of perfusion present while causing inflammation to the surrounding tissues 19.
The initiation of perfusion can occur once the cannulation is complete but before the tube coming from the LA is connected to the EVLP line. Perfusate should be run through to clear out any blood clots and this perfusate can empty into the chest wall without any issues. Switching the perfusate pump to manual mode and slowly increasing the flow rate to ~2 ml/min allows for close monitoring of the PA pressure. Pressures over 20-30 cmH2O can indicate an obstruction and watching for perfusate exiting the LA is also an indicator but this can be very hard to see. If the pressure does increase to over 20-30 cmH2O, stop the pump and recheck both cannulations. Once the pressure is constant around 10-20 cmH2O allow the perfusate to run through and into the chest cavity for 2 min. At this time the line from the LA can be connected to the EVLP circuit. The perfusate pump speed can be increased to 5-10 ml/min. As the fluid head progresses through the circuit, there will be an increase in the PA pressure due to the increase in height of the fluid head and therefore the static pressure. If the fluid cannot flow over the highest point in the line, it may be necessary to either apply a suction force on the opposite end of the line or attempt to lower the highest portion of the line. Once this issue is overcome, the perfusate should circulate without any issues.
A few issues should be monitored with respect to the ventilator. First, twisting of bronchi/trachea and heart-lung position may occur as the lungs become more edematous and the weight increases. It is important for the cannulae to remain in a relatively close anatomical position, therefore altering either or both cannulae may be necessary. Pressure or volume controlled ventilators as well as positive or negative ventilation can be used with this EVLP system. For the rat model, we have found using positive pressure, volume controlled ventilation works well at tidal volumes between 4-10 ml/kg and at positive end-expiratory pressures (PEEP) between 2-8 cmH2O. However, a PEEP of 8 cmH2O can cause a possible rupture at the bifurcation of the trachea. After each experiment (or set of experiments if performed back-to-back), the ventilation line leading to the trachea should be cleaned of any bronchoalveolar lavage (BAL) fluid that may have travelled up the trachea. This fluid will harden if left untouched and can completely block the ventilation line.
The perfusate composition is critical to a successful EVLP experiment. A 5% dextran mixture allows for lung perfusion that is close to physiologic conditions, maintains a stable oncotic pressure to drive fluid back into the vasculature to prevent edema and prevents thrombosis within the pulmonary vessels. It is important to note that some species of rats can be allergic to dextran which might cause pulmonary edema20. The contents of the perfusate was consistent across all experimental groups in this study, therefore the dextran content should not be a confounder. The oncotic pressure is a critical variable that has the potential to improve or produce tissue edema. Commercially available perfusion solutions that are optimized for cold static storage or normothermic perfusions have been used in this system to increase lung viability times. We note that some of these solutions contain albumin and one concern is the possibility of bovine albumin triggering an inflammatory response in the rodent lung. Although optimal perfusate composition is an ongoing subject of investigation, the perfusate needs to take into account the oncotic pressure, the osmotic pressure and buffering capacity. We recommend that the solution be based upon a modified Krebs–Henseleit solution or cell culture media. The oncotic pressure should be maintained by dextran or albumin, depending on the application. The perfusion pressure and flow rate affects the organ and supra-physiologic perfusion parameters can make the organ prone to mechanical trauma.
Visual Indicators during Experiment:
There are many visual cues as well as indications from real-time data that can be used to determine if an EVLP experiment is running well. The lung will remain the same size and will deflate to the same volume after every breath. There will also be no leaking from the lung itself. The PVR, lung weight, and compliance will remain relatively constant. Oxygen production will remain constant or increase slightly.
There are many visual indicators when the lung becomes compromised during an experiment. The lung becomes edematous and grows rapidly in size and weight. The color of the lung changes (from a tan-pink to white) and pockets of liquid can be identified in the tissue. If the trachea or lung ruptures from barotrauma or over distension, there will be bubbling from the point of injury (Figure 12B). Oxygen production will decrease and the PVR and compliance will dramatically increase as well.
The potential of using an EVLP model on small animals such as rodents opens the door for future studies improving the treatment of lung transplants. However, the small animal model requires a better understanding to truly mimic a lung transplantation. This model may be used in the future to improve medical treatments and define baseline parameters for future lung transplant studies.