1. Making the Standards: Serial Dilutions
2. Run the Samples for the Calibration Curve and the Unknown
3. Making the Calibration Curve
4. Results: Calibration Curve of Absorbance of Blue Dye #1

Figure 1. Calibration curves for UV-Vis absorbance of blue dye. Left: The absorbance was measured of different concentrations of blue dye #1. The responses level off after 10 µM, when the absorbance is over 1. The error bars are from repeated measurements of the same sample and are standard deviations. Right: The linear portion of the calibration curve is fit with a line, y=0.109*x + 0.0286. The unknown data is shown in black. Please click here to view a larger version of this figure.
Source: Laboratory of Dr. B. Jill Venton - University of Virginia
Calibration curves are used to understand the instrumental response to an analyte an…
Chapters in this video
0:00
Overview
0:51
Principles of Calibration Curves
2:53
Preparing the Standard Solutions
3:41
Preparing the Calibration Curve
5:26
Applications
7:19
Summary