In this study, we present a protocol for the differentiation of neural stem and progenitor cells (NPCs) solely induced by direct current (DC) pulse stimulation in a microfluidic system.
Physiological electric fields (EF) play vital roles in cell migration, differentiation, division, and death. This paper describes a microfluidic cell culture system that was used for a long-term cell differentiation study using microscopy. The microfluidic system consists of the following major components: an optically transparent electrotactic chip, a transparent indium-tin-oxide (ITO) heater, a culture media-filling pump, an electrical power supply, a high-frequency power amplifier, an EF multiplexer, a programmable X-Y-Z motorized stage, and an inverted phase-contrast microscope equipped with a digital camera. The microfluidic system is beneficial in simplifying the overall experimental setup and, in turn, the reagent and sample consumption. This work involves the differentiation of neural stem and progenitor cells (NPCs) induced by direct current (DC) pulse stimulation. In the stem cell maintenance medium, the mouse NPCs (mNPCs) differentiated into neurons, astrocytes, and oligodendrocytes after the DC pulse stimulation. The results suggest that simple DC pulse treatment could control the fate of mNPCs and could be used to develop therapeutic strategies for nervous system disorders. The system can be used for cell culture in multiple channels, for long-term EF stimulation, for cell morphological observation, and for automatic time-lapse image acquisition. This microfluidic system not only shortens the required experimental time, but also increases the accuracy of control on the microenvironment.
Neural precursor cells (NPCs, also known as neural stem and progenitor cells) can be as a promising candidate for neurodegenerative therapeutic strategy1. The undifferentiated NPCs have self-renewal capacity, multi-potency, and proliferative ability2,3. A previous study has reported that the extracellular matrix and molecular mediators regulate differentiation of NPC. The epidermal growth factor (EGF) and basic fibroblast growth factor (bFGF) promote NPC proliferation, thus maintaining the undifferentiated state4.
Previous studies have reported that electrical stimulation can regulate cell physiologic activities such as division5, migration6,7,8, differentiation1,9,10, and cell death11. Electric fields (EFs) play vital roles in the development and regeneration of the central nervous system development12,13,14. From 2009 to 2019, this laboratory has investigated cellular responses to the application of EF in the microfluidic system1,6,7,8,15,16,17. A multichannel, optically transparent, electrotactic (MOE) chip was designed to be suitable for immunofluorescence staining for confocal microscopy. The chip had high optical transparency and good durability and allowed the simultaneous conduct of three independent stimulation experiments and several immunostained conditions in a single study. The microfluidic system is beneficial in simplifying the overall experimental setup and, in turn, the reagent and sample consumption. This paper describes the development of a microfluidic cell culture system that was used for a long-term cell differentiation study.
1. Design and fabrication of the MOE chip
2. Coating poly-L-lysine (PLL) on the substrate in the cell culture regions
3. Preparation of the salt bridge network
4. Preparation of mNPCs
5. Setup of the microfluidic system for DC pulse stimulation (Figure 6)
6. Immunofluorescence assays of mNPCs after pulsed DC stimulation
NOTE: In this step, all reagent is pumped via the medium inlet using a syringe pump.
7. Image analysis and data processing
The detailed configuration of the MOE chip is shown in Figure 1. The microfluidic chip provides a beneficial approach for reducing the experimental setup size, sample volume, and reagent volume. The MOE chip was designed to perform three independent EF stimulation experiments and several immunostaining conditions simultaneously in a single study (Figure 3). In addition, the MOE chip, which has a high optical transparency is suitable for confocal microscopy examinations. The MOE chip is also designed to investigate the effects of different cell culture conditions (e.g., multiple EF stimulation, several drugs, different coating substrate, multiple series of cells) simultaneously in a single experiment.
The mNPCs were exposed to square-wave DC pulses (magnitude 300 mV/mm at a frequency of 100 Hz). The DC pulse stimulation was conducted for 48 h. The differentiated cells were immunostained with Tuj1 (neuron-specific class III β-tubulin), glial fibrillary acidic protein (GFAP to identify astrocytes), and oligodendrocyte marker O4. After the DC pulse treatment, the mNPCs expressed significantly high numbers of neurons (Tuj1+ cells) at DIV 7. At DIV 3, astrocytes (GFAP+ cells) were present at relatively higher levels in the stimulation groups than in the control (CTL) group. Compared with the CTL group, oligodendrocytes (O4+ cells) were significantly higher in the stimulation group at DIV 7 and DIV 14 (Figure 7). These results show that the DC pulse stimulation resulted in mNPCs differentiating into neurons, astrocytes, and oligodendrocytes simultaneously in stem cell maintenance medium. These results suggest that the MOE microfluidic system is suitable for a long-term cell differentiation study by microscopy.
Figure 1: The detailed configuration of the multichannel optically transparent electrotactic chip. (A) Exploded view of the MOE chip assembly. The MOE chip consists of PMMA sheets (50 mm x 25 mm x 1 mm), double-sided tape (50 mm x 25 mm x 0.07 mm), adaptors (10 mm x 10 mm x 6 mm), optical grade PMMA sheet (50 mm x 75 mm x 3 mm), double-sided tape (24 mm x 60 mm x 0.07 mm), and a cover glass (24 mm × 60 mm). There are three cell culture chambers in the MOE chip. The MOE chip has connecting holes for the medium inlet/outlet and the agar salt bridges. Cells were cultured in the cell culture region (width 3 mm x length 42 mm x height 0.07 mm). Figure 1A has been modified from Chang et al.6. (B) Photograph of the MOE chip comprising adaptors, PMMA sheets, double-sided tape, and cover glass. Abbreviations: MOE= multichannel optically transparent electrotactic; PMMA = polymethyl methacrylate. Please click here to view a larger version of this figure.
Figure 2: The fabrication and assembling processes of the MOE chip. (A) The designed patterns of the PMMA sheets or double-sided tape were fabricated using laser micromachining. (B) The individual PMMA sheets were cut by a CO2 laser scriber. (C) The multiple layers of the cleaned PMMA sheets were bonded together by a thermal bonder. Abbreviations: MOE= multichannel optically transparent electrotactic; PMMA = polymethyl methacrylate; CO2 = carbon dioxide. Please click here to view a larger version of this figure.
Figure 3: A photograph of the MOE chip. This figure has been modified from Chang et al.6. Abbreviation: MOE= multichannel optically transparent electrotactic. Please click here to view a larger version of this figure.
Figure 4: Medium and electrical connection to the MOE chip. (A) Photograph of the components for the medium flow network and the EF network in the MOE microfluidic system, including the PTFE tube, flat-bottom connector, cone connector, cone-Luer adaptor, white finger-tight plug, Luer adaptor, Luer lock syringe, black rubber bung, and the Ag/AgCl electrodes. (B) Photograph of the configuration for the medium flow network. Abbreviations: MOE= multichannel optically transparent electrotactic; EF = electric field; PTFE = polytetrafluoroethylene; Ag = silver; AgCl = silver chloride. Please click here to view a larger version of this figure.
Figure 5: A photograph showing the MOE chip on a microscope. Abbreviations: MOE= multichannel optically transparent electrotactic; Ag = silver; AgCl = silver chloride; ITO = indium-tin-oxide. Please click here to view a larger version of this figure.
Figure 6: The configuration and the system used for the DC pulse stimulation. (A) The configuration of the entire system for the DC pulse stimulation. The syringes connected to the MOE chip were used for medium infusion and waste efflux. The DC pulse in the chip was provided by a power supply conducted through the Ag/AgCl electrodes. The device setup was installed on the X-Y-Z motorized stage of an inverted phase-contrast microscope equipped with a digital camera. (B) A photograph showing the setup on a laboratory bench. Abbreviations: MOE= multichannel optically transparent electrotactic; Ag = silver; AgCl = silver chloride; ITO = indium-tin-oxide; EF = electric field. Please click here to view a larger version of this figure.
Figure 7: Differentiation of the mNPC cells in the control group (CTL) and in the DC pulse stimulation group at DIV 3, 7, and 14. The percentage of neuron (Tuj1+ cells), astrocytes (GFAP+ cells), and oligodendrocytes (O4+ cells) in (A–C) the CTL group and (D–F) in the stimulation (DC pulses) group. This figure has been published by Chang et al.1. Abbreviations: CTL: control; DC = direct current; Tuj1 = neuron-specific class III β-tubulin; GFAP = glial fibrillary acidic protein; O4 = oligodendrocyte marker O4. Please click here to view a larger version of this figure.
During the fabrication of the MOE chip, the adaptors are attached to the Layer 1 of the MOE chip with fast-acting cyanoacrylate glue. The glue is applied to 4 corners of the adaptors, and then pressure is applied evenly over the adaptors. Excess amount of glue must be avoided to ensure complete polymerization of the glue. Moreover, the completed MOE chip assembly is incubated in a vacuum chamber. This step helps to remove the bubbles between the PMMA layer, the double-sided tape, and the cover glass.
The choice of the electrode material is based on the fact that chloride ions, which are abundantly present in the medium, are the electrolytic products flowing through the cell culture region. During the EF stimulation experiment, the pH around the electrodes remained constant. A simpler configuration using platinum (Pt) as the electrode material electrolyzes water and generates hydrogen ions (H+) and hydroxide ions (OH–) at the positive electrode and the negative electrode, respectively, inducing pH changes in the culture region. Avoiding the use of Pt electrodes circumvents the problem of pH changes during the EF stimulation experiment.
The hot agarose and bubble-free agarose are essential during the preparation of the salt bridge network. The hot agarose has high fluidity and can be easily injected into the salt bridge network. Connect the Luer lock syringe to the Luer adaptor after injecting the 3% hot agarose into the Luer adaptor. During this step, the agarose will be pushed up into the Luer lock syringe so that a bubble-free firm connection of the salt bridge network can be achieved. Bubbles in the salt bridges increase the electrical resistance and hence, the anticipated electric current cannot be reached. After the agarose injection, it is important to wait for the agarose to cool down and solidify at room temperature for 10-20 min to prevent the formation of solidified agarose debris in the cell culture region.
The MOE chip is placed onto an ITO heater that is locked on a programmable X-Y-Z motorized stage. The entire system is built onto an inverted phase-contrast microscope equipped with a digital camera to monitor cell differentiation within the cell culture regions in the chip. It is convenient to observe the cell morphology and acquisition of the automatic time-lapse images in the MOE microfluidic system outside an incubator. This microfluidic system not only shortens the required experimental time, but also increases the accuracy of control on the microenvironment.
The mNPC cells grow as a suspension in culture media. However, mNPCs adhering to the PLL-coated plate in the MOE chip are critical for differentiation. Neurospheres formed by 30-40 cells are preferred for initiating mNPC differentiation. Overgrowth of mNPCs will impair cell survival during the differentiation process. Furthermore, after the pulsed DC stimulation, the immunofluorescence staining experimental can be affected by the flow rate. Hence, use several flow rates for different steps to avoid detaching cells during the wash.
In this study, a limitation of this technique is that the MOE chip cannot be reused because of the difficulty in thorough cleaning of the chip. However, the MOE chip can be placed under a phase-contrast microscope or a scanning confocal microscope directly. The water-tight design of the reported microfluidic system ensures that buffer/medium evaporation does not occur, maintaining the accurate concentration of the buffer/medium and the corresponding electrical properties. By reducing reagent volumes and the corresponding operation time, the MOE microfluidic system provides an efficient approach for studying cell differentiation.
A previous study has shown that EGF and bFGF promote NPC survival, expansion, and maintenance in the undifferentiated state4. In this study, the DC pulses induced the differentiation of the mNPCs in the stem cell maintenance medium that contained EGF and bFGF. Previous studies have reported that EF promotes differentiation of NPCs into neurons and/or astrocytes in differentiation medium without EGF and bFGF14,21,22. These results show that the mNPCs differentiated into neurons, astrocytes, and oligodendrocytes after the DC pulse stimulation. They also suggest that simple DC pulse treatment could control the fate of NPCs. With further optimization on the stimulation time, EF strength, or duty cycle, DC pulses may be applied to manipulate NPC differentiation and may be used for the development of therapeutic strategies that employ NPCs to treat nervous system disorders.
The authors have nothing to disclose.
The authors thank Professor Tang K. Tang, Institute of Biomedical Sciences, Academia Sinica, for his assistance in providing mouse neural stem and progenitor cells (mNPCs). The authors also thank Professor Tang K. Tang and Ms. Ying-Shan Lee, for their valuable discussion on the differentiation of mNPCs.
1 mm PMMA substrates (Layers 1-3) | BHT | K2R20 | Polymethyl methacrylate (PMMA), http://www.bothharvest.com/zh-tw/product-421076/Optical-PMMA-Non-Coated-BHT-K2Rxx-xx=-thickness-choices.html |
15 mL plastic tube | Protech Technology Enterprise Co., Ltd | CT-15-PL-TW | Conical bottomed tube with cap, assembled, presterilized |
3 mL syringe | TERUMO | DVR-3413 | 3 mL oral syringes, without needle |
3 mm optical grade PMMA (Layer 5) | CHI MEI Corporation | ACRYPOLY PMMA Sheet | Optical grade PMMA |
3-way stopcock | NIPRO | NCN-3L | Sterile disposable 3-way stopcock |
5 mL syringe | TERUMO | DVR-3410 | 5 mL oral syringes, without needle |
Adaptor | Dong Zhong Co., Ltd. | Customized | PMMA adaptor |
Agarose | Sigma-Aldrich | A9414 | Agarose, low gelling temperature |
Amplifier | A.A. Lab Systems Ltd | A-304 | High voltage amplifier |
AutoCAD software | Autodesk | Educational Version | Drafting |
B-27 supplement | Gibco | 12587-010 | B-27 supplement (50x), minus vitamin A |
Basic fibroblast growth factor (bFGF) | Peprotech | AF-100-18B | Also called recombinant human FGF-basic |
Black rubber bung | TERUMO | DVR-3413 | From 3 mL oral syringes, without needle |
Bovine serum albumin (BSA) | Sigma-Aldrich | B4287 | Blocking reagent |
Centrifuge | HSIANGTAI | CV2060 | Centrifuge |
CO2 laser scriber | Laser Tools and Technics Corp. | ILS-II | Purchased from http://www.lttcorp.com/index.htm |
Cone connector | IDEX Health & Science | F-120X | One-piece fingertight 10-32 coned, for 1/16" OD natural |
Cone-Luer adaptor | IDEX Health & Science | P-659 | Luer Adapter 10-32 Female to Female Luer, PEEK |
Confocal fluorescence microscope | Leica Microsystems | TCS SP5 | Leica TCS SP5 user manual, http://www3.unifr.ch/bioimage/wp-content/uploads/2013/10/User-Manual_TCS_SP5_V02_EN.pdf |
Digital camera | OLYMPUS | E-330 | Automatic time-lapse image acquisition |
Digital oscilloscope | Tektronix | TDS2024 | Measure voltage or current signals over time in an electronic circuit or component to display amplitude and frequency. |
Double-sided tape | 3M | PET 8018 | Purchased from http://en.thd.com.tw/ |
Dulbecco’s modified Eagle’s medium/Ham's nutrient mixture F-12 (DMEM/F12) | Gibco | 12400024 | DMEM/F-12, powder, HEPES |
Dulbecco's phosphate-buffered saline (DPBS) | Gibco | 21600010 | DPBS, powder, no calcium, no magnesium |
EF multiplexer | Asiatic Sky Co., Ltd. | Customized | Monitor and control the electric current in individual channels |
Epidermal growth factor (EGF) | Peprotech | AF-100-15 | Also called recombinant human EGF |
Fast-acting cyanoacrylate glue | 3M | 7004T | Strength instant adhesive (liquid) |
Flat bottom connector | IDEX Health & Science | P-206 | Flangeless male nut Delrin, 1/4-28 flat-bottom, for 1/16" OD blue |
Function generator | Agilent Technologies | 33120A | High-performance 15 MHz synthesized function generator with built-in arbitrary waveform capability |
Goat anti-mouse IgG H&L (Alexa Fluor 488) | Abcam | ab150117 | Goat anti-mouse IgG H&L (Alexa Fluor 488) preadsorbed |
Goat anti-rabbit IgG H&L (Alexa Fluor 555) | Abcam | ab150086 | Goat polyclonal secondary antibody to rabbit IgG – H&L (Alexa Fluor 555), preadsorbed |
Hoechst 33342 | Invitrogen | H3570 | Nuclear staining |
ImageJ software | National Institutes of Health | 1.48v | Analyze the fluorescent images |
Indium–tin–oxide (ITO) glass | Merck | 300739 | For ITO heater |
Inverted phase contrast microscope | OLYMPUS | CKX41 | For cell morphology observation |
K-type thermocouple | Tecpel | TPK-02A | Temperature thermocouples |
Luer adapter | IDEX Health & Science | P618-01 | Luer adapter female Luer to 1/4-28 male polypropylene |
Luer lock syringe | TERUMO | DVR-3413 | For agar salt bridges |
Mouse anti-GFAP | eBioscience | 14-9892 | Astrocytes marker |
Oligodendrocyte marker O4 antibody | R&D Systems | MAB1326 | Oligodendrocytes marker |
Paraformaldehyde (PFA) | Sigma-Aldrich | P6148 | Fixing agent |
Phosphate buffered saline (PBS) | Basic Life | BL2651 | Washing solution |
Poly-L-Lysine (PLL) | SIGMA | P4707 | Coating solution |
Precision cover glasses thickness No. 1.5H | MARIENFELD | 107242 | https://www.marienfeld-superior.com/precision-cover-glasses-thickness-no-1-5h-tol-5-m.html |
Programmable X-Y-Z motorised stage | Tanlian Inc | Customized | Purchased from http://www.tanlian.tw/ndex.files/motort.htm |
Proportional–integral–derivative (PID) controller | Toho Electronics | TTM-J4-R-AB | Temperature controller |
PTFE tube | Professional Plastics Inc. Taiwan Branch | Outer diameter 1/16 Inches | White translucent PTFE tubing |
Rabbit anti-Tuj1 | Abcam | ab18207 | Neuron marker |
Syringe pump | New Era Systems Inc | NE-1000 | NE-1000 programmable single syringe pump |
TFD4 detergent | FRANKLAB | TFD4 | Cover glass cleaner |
Thermal bonder | Kuan-MIN Tech Co. | Customized | Purchased from http://kmtco.com.tw/ |
Triton X-100 | Sigma-Aldrich | T8787 | Permeabilized solution |
Ultrasonic cleaner | LEO | LEO-300S | Ultrasonic steri-cleaner |
Vacuum chamber | DENG YNG INSTRUMENTS CO., Ltd. | DOV-30 | Vacuum drying oven |
White fingertight plug | IDEX Health & Science | P-316 | 1/4-28 Flat-Bottom, https://www.idex-hs.com/store/fluidics/fluidic-connections/plug-teflonr-pfa-1-4-28-flat-bottom.html |