Matrix-assisted laser desorption ionization (MALDI) is a mass spectrometry ion source ideal for the analysis of biomolecules. Instead of ionizing comp…
Matrix-assisted laser desorption ionization (MALDI) is a mass spectrometry ion source ideal for the analysis of biomolecules. Instead of ionizing compounds in the gaseous state, samples are embedded in a matrix, which is struck by a laser. The matrix absorbs the majority of the energy; some of this energy is then transferred to the sample, which ionizes as a result. Sample ions can then be identified using a time-of-flight analyzer (TOF).
This video covers principles of MALDI-TOF, including matrix selection and how TOF is used to elucidate mass-to-charge ratios. This procedure shows the preparation of a MALDI plate, the loading of samples onto the plate, and the operation of the TOF-mass spectrometer. In the final section, applications and variations are shown, including whole-cell analysis, characterization of complex biological samples, and electron spray ionization.
Matrix-assisted laser desorption ionization, or MALDI, is a mass spectrometry ion source ideal for the analysis of biomolecules. Most ion sources remove structural information from large, fragile biomolecules. MALDI maintains structural integrity, and therefore information, while accelerating the molecules into the mass analyzer, which separates the compounds based on size and charge. The most commonly coupled with MALDI is the time of flight, or TOF, mass analyzer. This video will show the concepts of MALDI ionization, a general procedure, and some of its uses in biochemistry.
For mass spectrometry to function, molecules must be ionized into the gaseous state. In MALDI, the sample is embedded in a matrix, typically an organic compound containing aromatic and conjugated double bonds.
When a laser pulse strikes this mixture the matrix absorbs the majority of the energy, rapidly heats, and is desorbed, or released, from the surface. The energized matrix transfers some of its energy to the biomolecules, desorbing and then ionizing them.
MALDI is typically paired with a time of flight, or TOF, mass analyzer. An electric field applies kinetic energy to the ions, moving them into a field-free region called a drift tube. The velocity of the ions as they move through the tube is related to their mass-to-charge ratio, so heavier particles travel slower through instrument. A detector at the end of the tube measures each ion's flight time. With this knowledge, as well as the tube length and applied field strength, the mass-to-charge ratio of each ion can be elucidated.
This plot of signal intensity to mass-to-charge-ratio, known as a mass spectrum, can be compared to a library of collected spectra. If no matches are found, it can molecules can be identified by further techniques, such as tandem mass spectrometry. For more information, see this collection's video on the topic.
Now that the basics of MALDI-TOF have been discussed, let's look at the process in the laboratory.
Before beginning an experiment, it's important to consider the choice of matrix from which samples will be desorbed. It must absorb the laser energy, be stable in a vacuum, not react with the target molecules, and be able to desorb. Depending on the sample, different matrices are preferred. For a large protein, a combination of CHCA and DHB has shown better separation of the peaks, called resolution, than the individual matrices.
There are a number of ways to prepare samples. We'll show what is known as the "double-layer", or "sandwich," method. To begin, clean the MALDI plate with ultra-pure reagents, as mass spectrometry is very sensitive to contamination. Dry the plate with a stream of inert gas.
Next, a saturated matrix solution is made, typically with an organic solvent . The solution is streaked onto the MALDI plate and dried. A second saturated solution of matrix containing trifluoroacetic acid, or TFA, is prepared. TFA helps ions into the gaseous phase.
Next, the sample solution is added on top of the dried matrix spot. Add the matrix solution containing TFA on top of the sample, thereby completing the matrix "sandwich". Homogeneity of the spot can be verified under a low-powered microscope.
Plate a calibration standard, which is a mixture with a wide range of known masses and is used to correlate the time-of-flight to m/z. Finally, plate the matrix alone as a negative control.
To analyze the spots, place the target plate into the instrument. Ensure there's no debris present, allowing for the formation of a tight vacuum. In the software, select the standard, negative control, and samples of interest. Label the spots with the correct identification.
The ion source and lens voltages can be manipulated to improve performance of the analysis. This will depend on the specifics of the instrument and sample. Focus on the standard spot and calibrate the instrument with the software.
Next, collect spectra from each of the sample spots. Try a few different locations on the spot to maximize the quality of the collected data. Once finished, the MALDI plate can be collected and reused after cleaning.
Now that we've reviewed a procedure, let's look at some of the ways MALDI is utilized, and a different ionization technique.
In addition to biomolecules, MALDI can be used to analyze living cells. Macrophages are immune cells that take on one of several different forms, based on their microenvironment. After exposing the cells to various signaling molecules, or cytokines, they can be added directly to the plate, and analyzed. The MALDI spectra can be used as unique "fingerprints", depending on the cytokine used.
Complex biological samples like mammalian sebaceous secretions require a step of purification before MALDI analysis. Thin layer chromatography is one such technique that relies on the components' polarity. The compounds are collected from the TLC pate, purified, and transferred to a MALDI matrix. The resulting spectra verify the identity and purity of the separated biomolecules from the mammalian sebaceous secretions.
Another common ion source for biomolecules is electrospray ionization, or ESI. In this method, the sample is injected into the instrument, where a high voltage is applied, creating an aerosol of charged droplets. As the solvent in the droplet evaporates, the charge is moved to the sample molecules, till they are completely gaseous. ESI doesn't require the spotting procedure, and the sample can be injected directly into the instrument. On the other hand, ESI is more sensitive to the presence of buffer components and other contaminants, meaning MALDI is more robust.
You've just watched JoVE's video on MALDI mass spectrometry. This video described the theory behind the instrument, went over a general procedure, and covered some of the uses of the technique. Thanks for watching!
View the full transcript and gain access to JoVE Science Education videos
Q1: How does MALDI preserve structural information in biomolecules during ionization?
MALDI maintains structural integrity by embedding samples in a matrix that absorbs laser energy instead of directly ionizing the biomolecules. When the laser strikes the mixture, the matrix rapidly heats and desorbs, transferring energy to the biomolecules to ionize them gently. This preserves critical structural information that other ion sources remove from large, fragile molecules.
Q2: What role does the matrix play in MALDI ionization?
The matrix is an organic compound containing aromatic and conjugated double bonds that absorbs the majority of laser energy. When struck by the laser pulse, it rapidly heats and desorbs from the surface, transferring some energy to embedded biomolecules. The matrix must absorb laser energy, remain stable in vacuum, not react with target molecules, and be able to desorb effectively.
Q3: How does a time-of-flight analyzer determine the mass-to-charge ratio of ions?
An electric field accelerates ions into a field-free drift tube, where their velocity correlates with mass-to-charge ratio—heavier particles travel slower. A detector measures each ion's flight time through the tube. Using the tube length, applied field strength, and flight time, the mass-to-charge ratio of each ion can be calculated and plotted as a mass spectrum.
Q4: What is the double-layer or sandwich method for MALDI sample preparation?
The sandwich method involves streaking a saturated matrix solution onto a clean MALDI plate and drying it. Next, sample solution is added on top of the dried matrix spot. Finally, a second saturated matrix solution containing trifluoroacetic acid is applied on top, creating a three-layer sandwich. This method improves sample homogeneity and ionization efficiency.
Q5: Why is MALDI more robust than electrospray ionization for complex biological samples?
MALDI is more robust because it is less sensitive to buffer components and contaminants present in samples. Electrospray ionization, by contrast, requires sample injection and is highly susceptible to contamination, making it less reliable for complex biological samples. MALDI's spotting procedure and matrix-based approach provide better tolerance for impurities.
Q6: How can MALDI be used to analyze living cells and generate cellular fingerprints?
MALDI can directly analyze living cells like macrophages after exposing them to signaling molecules called cytokines. Cells are added directly to the MALDI plate and analyzed. The resulting MALDI spectra serve as unique cellular fingerprints that vary depending on the cytokine used, allowing researchers to distinguish different cell states based on their molecular composition.
Q7: What preliminary step is required before MALDI analysis of complex biological samples?
Complex biological samples like mammalian sebaceous secretions require purification before MALDI analysis. Thin layer chromatography is one technique that separates components based on polarity. Compounds are collected from the chromatography plate, purified, and transferred to a MALDI matrix. The resulting spectra verify the identity and purity of separated biomolecules, and unidentified compounds can be analyzed using tandem mass spectrometry principle instrumentation uses.