The in vivo application of implants and biomaterials, which interact with human blood, requires intense preclinical testing focusing on the investigation of various markers of the hemostatic system. The International Organization for Standardization 10993-4 (ISO 10993-4) specifies the central principles for the evaluation of blood-contacting devices (i.e., stents and vascular grafts) and considers the device design, clinical utility, and materials needed1.
Human blood is a fluid that contains various plasma proteins and cells, including leukocytes (white blood cells [WBCs]), erythrocytes (red blood cells [RBCs]), and platelets, which carry out complex functions in the human body2. The direct contact of foreign materials with blood can cause adverse effects, such as activation of the immune or coagulation system, which can lead to inflammation or thrombotic complications and serious issues after implantation3,4,5. Therefore, in vitro hemocompatibility validation offers an opportunity prior to implantation to detect and exclude any hematologic complications that may be induced upon contact of the blood with a foreign surface6.
The presented flow loop model was established to assess the hemocompatibility of neurovascular stents and similar devices by applying a flow rate of 150 mL/min in tubing (diameter of 3.2 mm) to mimic cerebral flow conditions and artery diameters2,7. Besides the need for an optimal in vitro model, the source of blood is an important factor in gaining reliable and unaltered results when analyzing hemocompatibility of a biomaterial8. The collected blood should be used immediately after sampling to prevent changes caused by prolonged storage. In general, a gentle collection of blood without stasis using a 21 G needle should be performed to minimize the preactivation of platelets and the coagulation cascade during blood drawing. Furthermore, donor exclusion criteria include those who smoke, are pregnant, are in a poor state of health, or have taken oral contraceptives or painkillers during the previous 14 days.
This study describes an in vitro model for the extensive hemocompatibility testing of stent implants under flow conditions. When comparing uncoated to fibrin-heparin-coated stents, results of the comprehensive hemocompatibility tests reflect improved hemocompatibility of the fibrin-heparin-coated stents9. In contrast, the uncoated stents induce activation of the coagulation cascade, as demonstrated by an increase in thombin-antithrombin III (TAT) concentrations and loss of blood platelet numbers due to the adhesion of platelets to stent surface. Overall, integrating this hemocompatibility model as a preclinical test is recommended to detect any adverse effects on the hemostatic system that are caused by the device.