A major source of measurement error arises from the capillaries that have defects either from the cutting process (Figure 5A,B) or the coating process (Figure 5D). Both types of defects lead to errors in determining the bubble shape and size by the optical image analysis system, leading to inaccurate surface tension values. It is important to carefully examine each new capillary after it is pulled and coated under the optical microscope before inserting the capillary into the CPM. A miscut capillary must be discarded, but a poorly coated capillary can be acid-cleaned and re-coated to improve the bubble pinning at the end of the capillary (step 2 of the Protocol). Capillaries work best if the end cut is perfectly perpendicular to the capillary (Figure 5C) and the bubble pins directly at the end of the capillary (Figure 5E). The hydrophobic coating on the capillary will become less effective at pinning with use, requiring the capillary to be re-cleaned and re-coated.
Representative data for surfactant adsorption vs. time is presented in Figure 6. Previous experimental techniques such as a pendant or sessile drops used to measure surfactant adsorption did not have a mechanism to dynamically adjust the capillary pressure as the change in surface tension causes the bubble area to change during adsorption30,36,37. In fact, for larger bubbles and drops, changes in the bubble or drop shape (and hence surface area) are required to determine the surface tension from the analysis of the interface shape as the capillary pressure is not measured independently and the capillary pressure varies over the drop or bubble surface37. This also complicates the analysis of the adsorption because as surfactant adsorbs to the interface, the surface tension decreases, and to satisfy the Laplace equation the surface area of the bubble must increase, requiring additional surfactant to adsorb to reach equilibrium. In the CPM, a fixed capillary pressure requires that the initial bubble radius must be within a small range prior to surfactant adsorption to prevent the bubble from ejecting from the capillary if the surface tension decreases too much. Surfactant adsorption dynamics are often modeled by the classic Ward-Tordai equation31, which describes the adsorption of surfactant molecules to a clean interface of constant interfacial area. While the Ward-Tordai equation can be modified to account for the changing surface area, this introduces additional parameters and greatly complicates the analysis38,39.
To overcome these issues, a model-based feedback loop was developed using the Laplace equation that holds the curvature (and surface area) of the bubble constant throughout the adsorption process by dynamically adjusting the capillary pressure. There are significant differences in the rate of change of the surface tension because the area of the bubble is not constantly increasing. The changes in the bubble area during adsorption are not constant with time as the surface tension changes slowly at first, and then rapidly accelerates prior to equilibration. An additional complication is that the fractional change in the area depends on the initial bubble radius. An additional benefit of constant bubble radius is that imaging the interface is simplified as the bubble surface remains fixed, which simplifies focusing the CFM. During the adsorption process, as surfactant adsorbs to the interface (Video 1), the fluorescent signal from the interface increases. If the surfactant forms surface domains, these domains can be observed forming and growing22.
The changes in surface tension during area oscillations are shown in Figure 7. In previous versions of the CPM, oscillations were made in the bubble capillary pressure; however, generating a sine wave in capillary pressure does not translate directly into a sine wave in surface area as the two are related via the Laplace equation. By taking advantage of a model-based feedback loop using the Laplace equation, oscillations are created in the area rather than in capillary pressure, leading to data that is easier to analyze and collect over a larger range of amplitudes. As a result, the surface tension vs. area data collected from this method can be used to directly calculate the interfacial dilatational modulus of the surfactant layer:
(Figure 8), where
is the total stress of the system and τstress is non-isotropic deviatoric stress often absent in simple surfactant solutions4,33. Thus, for a simple surfactant system,
. For interfaces where elastic networks can be formed, such as surface-active proteins, extra stresses are often present and thus must be accounted for when defining the dilatational modulus. Video 2 shows a CFM video of the motion of black LC domains in a continuous colored LE phase matrix in phospholipid monolayers. The distinct LC domains on the interface reorganize into a branching network that covers the interface when oscillations take place on the curved bubble22,40. The Other Area Oscillations tab can be used to create sawtooth, square, and triangular waves as seen in Supplemental Figure 3 and the Compression tab allows for constant rate area compression and expansion.
For solvent exchange studies, a surfactant is first allowed to adsorb to the interface, and then the reservoir liquid is exchanged to allow a second surface-active species to contact that interface. It is possible to examine the change in surface tension as the second surfactant competes with the original surfactant at the interface. The surface dilatational modulus is often a more sensitive probe of the surfactant exchange along with the surface morphology via CFM. Figure 9 shows the change in surface tension, surface dilatational modulus, and surface morphology as one such solvent exchange takes place. While the specifics of such an exchange can vary, a change to any of the three properties could indicate integration of the second component into the monolayer or solvation of the primary component into the bulk. A second fluorescent tag could be attached to the secondary species to observe its interaction with the interface from the CFM images.

Figure 1: Capillary Treatment. (A) Image showing the scoring of the capillary. The glass scoring ceramic is held in a clamp to hold it steady. (B) Acid-cleaning of the capillary. The acid cleaning solution is pulled into the capillary with the vacuum pump. (C) Hydrophobization of the capillary. Silane solution plug held inside the capillary Please click here to view a larger version of this figure.

Figure 2: Cell construction. (1) Large aluminum cell holder, (2) Fluoroelastomer gasket (four in total), (3) glass slide (two in total), (4) PEEK cell, and (5) small aluminum cell holder. When assembled, a fluoroelastomer gasket is placed on either side of each glass slide. The cell is held together with screws and bolts. The zoomed-in image of the PEEK cell shows the locations of the various ports: (6) capillary port, (7) solvent exchange inlet, (8) solvent exchange outlet, and (9,10) temperature control jacket inlet and outlet. A PEEK plug can be used to attach the tubing or capillary to the cell. Ports that are not being used can be completely closed by plugs without channels. Please click here to view a larger version of this figure.

Figure 3: Schematic of CPM/CFM, not to scale. (1) the CPM cell, (2) the capillary tube with a bubble at the tip, (3) confocal microscope objective, (4) microscope camera objective with filter, (5) CPM light source, (6) microfluidic pump, (7) safety valve, (8) fluid exchange inlet, (9) fluid exchange outlet, (10) peristaltic pump, (11) exchange fluid reservoir, (12) fluid exchange waste, (13) direct to cell syringe, (14) temperature control jacket inlet and outlet, and (15) temperature-controlled reservoir and pump. Please click here to view a larger version of this figure.

Figure 4: CPM virtual interface. (1) the file path where the data will be saved; (2) system parameters, comments, and the Save button. All fields in this area are saved to the final data file; (3) the CPM camera image; (4) settings controlling the image analysis, annulus measurement, bubble reset, and frames per second tracking; (5) the Bubble Reset button; (6) the Collect Data button, data recording rate control, and data collection indicators; (7) controls for all operating mode centerline values, oscillation amplitude, and oscillation frequency; (8) operating mode switch: clicking each tab changes to that control mode. Each mode shows the pressure signal being sent to the pump in the "Pressure Signal" graph as well as some additional controls; (9) live surface tension data; (10) live pressure data; (11) live radius of curvature data; (12) live surface area data; and (13) live surface tension and surface area data, which can be used to roughly determine the phase angle during an oscillation study. Please click here to view a larger version of this figure.

Figure 5: Capillary Defects. (A) and (B) Miscut capillaries; (C) correctly cut capillary, (D) capillary with bad pinning due to poor or degraded coating, and (E) properly pinned capillary. The red arrows in D and E indicate where the bubbles are pinned. For the best results, the bubble will pin at the capillary tip. Please click here to view a larger version of this figure.

Figure 6: Adsorption study microtensiometer results for both constant pressure (orange) and constant area (blue) adsorptions. The bubble surface area for the constant area adsorption increases significantly throughout the study and causes the adsorption to take longer to reach the same surface tension. Please click here to view a larger version of this figure.

Figure 7: Typical surface area control oscillation. (A) Pressure, (B) curvature, and (C) surface area data. The surface area data is a sinusoid while the pressure and curvature data are not, as evidenced by the centerline values not being at the midpoint of the oscillation. The mathematical relationship between the three values means only one can be a true sinusoid. Please click here to view a larger version of this figure.

Figure 8: Sample rheological results after analysis. Dilatational modulus of Lyso PC (1-palmitoyl-2-hydroxy-sn-glycero-3-phosphocholine) as a function of frequency for increasing concentrations of Lyso PC for ~45 µm radius bubbles. Concentrations >0.1 mM of Lyso PC that accompany inflammation decrease the dilatational modulus over the range of normal ventilation/ breathing rates (yellow) to make 2ε−γ < 0, which is the crossover value for inducing the Laplace instability (dotted red line). Low concentrations of Lyso PC ≤0.01 mM, that may occur in normal lungs do not induce instability. At frequencies >10 rad/sec, all Lyso PC concentrations are above the crossover, and would not be susceptible to the Laplace instability. Solid red lines are fits of theory to the data. Figure reproduced from reference9. Please click here to view a larger version of this figure.

Figure 9: CFM and CPM results for a solvent exchange study for lung surfactant exchanged with DI water and then Lyso PC. (A) shows how surface tension and surface dilatational modulus change throughout the study. The graph is separated into four regions: when the lung surfactant is adsorbed to the interface (blue), when the LS is exchanged with DI water (green), when the exchange solution is switched to a Lyso PC solution (red), and when the cell is filled with the Lyso PC solution (orange). The properties can be seen to change throughout the various exchanges indicating the interface is changing. (B) shows a confocal image of the lung surfactant adsorbed to the interface before exchange and (C) shows the same surface after the exchange with the Lyso PC solution is complete. In both cases, the white dashed circle indicates the inner edge of the capillary. The structure of the domains on the monolayer changes drastically after the solvent exchange, corroborating the CPM results. Please click here to view a larger version of this figure.
Video 1: Confocal video of constant pressure adsorption study for lung surfactant. The false-color shows the distance in the z-direction with the color bar on the left side of the video, with purple indicating the bubble near the capillary and green being the top of the bubble. The interface is initially dimly illuminated as only a little of the fluorescent surfactant is adsorbed. As more and more surfactant adsorbs, the bubble starts to grow as the color shifts more to green and the interface becomes populated by black LC domains that can move across the interface. Aggregates of surfactant in the solution can be seen floating in the solution as bright amorphous shapes and several settle onto the bubble interface, disintegrating and depositing their surfactant onto the interface. Please click here to download this Video.
Video 2: Confocal video of oscillation study for lung surfactant. The false-color shows the distance in the z-direction with the color bar on the left side of the video. The surface is subjected to several different oscillation frequencies and the dark LC domains on the interface can be seen to change throughout the oscillations. Please click here to download this Video.
Supplementary Figure 1: Example of an intermediate step in the code to determine the dilatational rheology. When this screen appears, the user should left-click at the left-most edge of the oscillation to analyze, and then left-click the right-most edge. Multiple oscillations can be analyzed so that the user could left-click on 1, 2, 3, and 4, and then right-click to analyze those two oscillations. The oscillations shown are of different amplitudes and frequencies. Please click here to download this File.
Supplementary Figure 2: Example of the graphical results produced by the dilatational rheology code. This shows the fits of sinusoids to the oscillations in pressure, radius, surface area, and surface tension as well as the Fourier transform of each oscillation. Ideally, the second harmonic in the Fourier transform should be less than 10% of the first harmonic for the surface area and surface tension. Please click here to download this File.
Supplementary Figure 3: Alternate operation modes. (A) Sine wave, (B) Sawtooth wave, (C) Square wave, (D) Triangular wave, (E) Constant rate expansion, and (F) Constant rate compression. The compression and expansive modes allow for Langmuir type isotherms to be created for insoluble surfactants. Please click here to download this File.
Supplemental Coding File 1: Microtensiometer Virtual Interface.vi. Please click here to download this File.
Supplemental Coding File 2: Dilatational_Rheology_Analysis.m. Please click here to download this File.