Method Article

Monitoring Changes in the Intracellular Calcium Concentration and Synaptic Efficacy in the Mollusc Aplysia

DOI:

10.3791/3907

July 15th, 2012

In This Article

Summary

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

We demonstrate how changes in the intracellular free calcium concentration and synaptic efficacy can be simultaneously monitored in a ganglion preparation of Aplysia. We image intracellular calcium using a fluorescent dye, Calcium Orange, and induce and monitor synaptic transmission with sharp (intracellular) electrodes.

Abstract

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

It has been suggested that changes in intracellular calcium mediate the induction of a number of important forms of synaptic plasticity (e.g., homosynaptic facilitation) 1. These hypotheses can be tested by simultaneously monitoring changes in intracellular calcium and alterations in synaptic efficacy. We demonstrate how this can be accomplished by combining calcium imaging with intracellular recording techniques. Our experiments are conducted in a buccal ganglion of the mollusc Aplysia californica. This preparation has a number of experimentally advantageous features: Ganglia can be easily removed from Aplysia and experiments use adult neurons that make normal synaptic connections and have a normal ion channel distribution. Due to the low metabolic rate of the animal and the relatively low temperatures (14-16 °C) that are natural for Aplysia, preparations are stable for long periods of time.

To detect changes in intracellular free calcium we will use the cell impermeant version of Calcium Orange 2 which is easily 'loaded' into a neuron via iontophoresis. When this long wavelength fluorescent dye binds to calcium, fluorescence intensity increases. Calcium Orange has fast kinetic properties 3 and, unlike ratiometric dyes (e.g., Fura 2), requires no filter wheel for imaging. It is fairly photo stable and less phototoxic than other dyes (e.g., fluo-3) 2,4. Like all non-ratiometric dyes, Calcium Orange indicates relative changes in calcium concentration. But, because it is not possible to account for changes in dye concentration due to loading and diffusion, it can not be calibrated to provide absolute calcium concentrations.

An upright, fixed stage, compound microscope was used to image neurons with a CCD camera capable of recording around 30 frames per second. In Aplysia this temporal resolution is more than adequate to detect even a single spike induced alteration in the intracellular calcium concentration. Sharp electrodes are simultaneously used to induce and record synaptic transmission in identified pre- and postsynaptic neurons. At the conclusion of each trial, a custom script combines electrophysiology and imaging data. To ensure proper synchronization we use a light pulse from a LED mounted in the camera port of the microscope. Manipulation of presynaptic calcium levels (e.g. via intracellular EGTA injection) allows us to test specific hypotheses, concerning the role of intracellular calcium in mediating various forms of plasticity.

Protocol

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

1. Preparation

  1. Anesthetize the animal by injecting 75-100 ml isotonic magnesium chloride solution. The Aplysia we use for imaging are generally 150-200 grams and are obtained from Marinus Scientific.
  2. Pin the anesthetized animal to a wax covered dish. Syringe needles work well for this purpose; sterile techniques are not necessary. Using gross forceps and standard scissors make an incision in the animal's foot and expose the buccal mass. Locate the buccal ganglion. Using spring scissors and fine forceps, carefully free the ganglion by cutting all buccal nerves.
  3. Remove the ganglion and place it in a Sylgard coated dis....

Access restricted. Please log in or start a trial to view this content.

Discussion

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

We demonstrate techniques that can be used to simultaneously monitor the intracellular calcium concentration and evaluate the efficacy of synaptic transmission. These techniques are useful for determining how various forms of short-term plasticity are mediated.

The imaging is carried out with a fluorescence microscope and CCD camera. These equipment requirements are relatively modest when compared to most functional imaging set-ups. The technique is simple and easy to learn. While imaging wit.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

We have nothing to disclose.

Acknowledgements

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

A PHS Grant (MH51393) supported this work. Some of the Aplysia we use are provided by the National Resource for Aplysia of the University of Miami under Grant RR10294 from the National Center for Research Resources, NIH.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Calcium OrangeInvitrogenC-3013
EGTASigmaE-4378
Calcium calibration buffer kitInvitrogenC-3008MPUseful for testing the sensitivity and dynamic range of the signal

Table 1. Reagents used.

FN-1 upright fluorescence microscopeNikon Instrumentswith Narishige ITS-FN1 stage
NMN-21 manipulatorsNarishigemounted on stage with magnets
CoolSNAP HQ2 CCD cameraPhotometrics
NIS elements ARNikon Instrumentswe used version 3.22
10X/0.3w Plan Fluor objectiveNikon Instrumentsthis water immersion lens has a very long working distance of 3.5 mm
X-Cite 120 PC metal halide lampEXFOused for fluorescence imaging
LS-DWL halogen lampSumica
ET-CY3 filter setChroma Technology
Power 1401 A/D converterCambridge Electronic Designsampling was done at 3 kHz
Spike II SoftwareCambridge Electronic Designwe used version 7.07
SEC-10 LX amplifierNPI electronicsused with a 10X headstage
Model 410 amplifierBrownlee precisionused to amplify and filter the signal
WS-4minus k Technologyvibration isolation for imaging
cooling platformcustom madebrass plate through which ice water is pumped at a variable rate

Table 2. Equipment used.

References

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,
  1. Zucker, R. S., Regehr, W. G. Short-term synaptic plasticity. Annu. Rev. Physiol. 64, 355-405 (2002).
  2. Eberhard, M., Erne, P. Calcium binding to fluorescent calcium indicators: Calcium green, calcium orange and calcium crimson. Biochem. Biophysical Res.....

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Tags

Calcium ImagingIntracellular RecordingAplysia GanglionCalcium Orange DyeElectrophysiology ImagingFluorescent MicroscopyCCD CameraSynaptic Plasticity