Lipid peroxidation is a process in which free radicals, such as reactive oxygen species and reactive nitrogen species, attack carbon-carbon double bonds in lipids, a process that involves the abstraction of a hydrogen from a carbon and insertion of an oxygen molecule. This process leads to a mixture of complex products including, lipid peroxyl radicals, and hydroperoxides as the primary products, as well as malondialdehyde (MDA) and 4-hydroxynonenal as predominant secondary products1.
MDA has been widely used in biomedical research as a marker of lipid peroxidation due to its facile reaction with thiobarbituric acid (TBA). The reaction leads to the formation of MDA-TBA2, a conjugate that absorbs in the visible spectrum at 532 nm and produces a red-pink color2. Other molecules derived from lipid peroxidation besides MDA can also react with TBA and absorb light at 532 nm, contributing to the overall absorption signal that is measured. Similarly, MDA can react with most other major classes of biomolecules, potentially limiting its accessibility for reaction with TBA3,4. As such, this traditional assay is simply considered to measure “thiobarbituric acid reactive substances” or TBARS5.
When correctly applied and interpreted, the TBARS assay is generally considered a good indicator of the overall levels of oxidative stress in a biological sample6. Unfortunately, as documented by Khoubnasabjafari and others, the TBARS assay is often conducted and interpreted in ways that facilitate dubious conclusions3,4,7,8,9,10,11. The causes for this are rooted primarily in sample-related pre-analytical variables and a lack of assay ruggedness that prohibits seemingly minor variations in assay protocol without substantial changes in assay results1,7,12,13.
Preanalytical variables related to biospecimen handling and storage (e.g., blood plasma kept temporarily at -20 °C)14,15 can have a major impact on TBARS assay results16,17; so much so, that TBARS assay results should not be compared across different laboratories unless warranted by explicit interlaboratory analytical validation data. This recommendation is akin to how western blots are commonly used and interpreted. Comparisons of band densities are valid for within-blot and perhaps within-laboratory studies, but comparing band densities between laboratories is generally considered an invalid practice.
Some researchers have suggested that MDA as measured by the TBARS assay simply does not meet the analytical or clinical criteria required of an acceptable biomarker3,9,10,18,19. Indeed, if the assay had not been developed over 50 years ago, it probably would not have gained the widespread use and tacit acceptability that it has today. Although there are other assays with greater analytical sensitivity, specificity, and ruggedness used for determining oxidative stress, TBARS assay based on absorbance at 532 nm remains by far one of the most commonly used assays for the determination of lipid peroxidation20, and thereby assessment of oxidative stress.
The TBARS assay can only be found as an expensive kit (over 400 U.S. dollars), in which the instructions do not provide detailed information on most concentrations of the reagents used. Additionally, the reagents provided can only be used for one experiment, because only one colorimetric standard curve can be made per kit. This can be problematic for researchers who intend to determine levels of oxidation within a few samples at different timepoints, because the same standard curve cannot be used at multiple times. Hence, multiple kits need to be purchased for multiple experiments. Currently, unless an expensive kit is purchased, there is not a detailed protocol available for how to perform a TBARS assay. Some researchers in the past have vaguely described how to perform a TBARS assay21,22, but neither a fully detailed protocol or comprehensive video on how to conduct the TBARS assay without an expensive kit is available in the literature.
Here we report a detailed, analytically validated for-purpose methodology on how to perform a TBARS assay in a simple, reproducible, and inexpensive way. Changes in the lipid peroxidation of human serum, HepG2 lysates, and low density lipoproteins upon treatment with Cu(II) ions are demonstrated as illustrative applications for the TBARS assay. Results demonstrate that this TBARS assay is consistent and reproducible on a day-to-day basis.